AIRSEP CORPORATION · Model AS077-1
Portable Oxygen Concentrator, 1-5 LPM equivalent.
Intended use: A mobile/portable mains electricity (AC-powered) device designed to concentrate oxygen (O2) from ambient air and deliver the concentrated O2, typically through an attached nasal cannula, to a patient requiring oxygen therapy. It processes the air through an internal filtration system, e.g., a molecular sieve (zeolite granules or membranes), having a large total surface area to separate nitrogen (N2) from the air. It typically consists of an air compressor, filters, dual chambers, a reservoir and controls. The O2 concentration is variable depending on the flow rate utilized. It is used in a home, institution, or a vehicle and typically has internal batteries for transportable use.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Oxygen-purity alarm, delivered O2 below spec, oxygen indicator amber / red.
Root cause: Aged sieve beds / molecular sieve, a valve-block fault, an air leak, or an inlet-filter blockage.
Symptoms: Weak or no flow, compressor noisy or hot, thermal cut-out, unit shuts down.
Root cause: Worn compressor / valve heads, or a cooling-fan fault.
Symptoms: Short run time, will not reach full charge, sudden shutdown.
Root cause: Aged battery pack or a charger fault.
Symptoms: Not triggering on inhalation, weak or no bolus, breath-detect alarm.
Root cause: Pressure-sensor / trigger fault, a cannula leak, or a nasal-cannula mismatch.
Symptoms: Reduced output, unit hot, over-temperature shutdown.
Root cause: Dirty gross-particle / cabinet filter or a blocked air path.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$ | Easy |
| Sieve-bed / column assembly | Periodic replacement | $$$ | Moderate |
| Compressor / valve-head kit | Periodic replacement | $$$ | Difficult |
| Inlet / cabinet filters | Consumable | $ | Easy |
| Pressure / trigger sensor | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio