Air Techniques · Model Peri-Pro III
Process all intraoral film sizes 0, 1, 2, 3 and 4 Rollerless transport system Slim space-saving design Easier and faster maintenance, ready to read in only 5 minutes External water drain valve in the base and a refill access door built right into the cover Peri-Pro III has an electronically controlled heating system to maintain an optimum 75° F chemistry temperature Film surfaces are never touched Up to 8 films may be loaded at a time.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Fogged, streaked, mottled or under-developed films; density drift on the sensitometric strip.
Root cause: Exhausted or wrongly mixed chemistry, wrong developer temperature, or dirty / worn rollers.
Symptoms: Film jams between racks, does not advance, or comes out crumpled.
Root cause: Worn or misaligned transport rollers / racks, worn drive gears, or a crossover-rack fault.
Symptoms: Developer not at set temperature, density varies through the day.
Root cause: Heater, thermostat / sensor, or recirculation-pump fault.
Symptoms: Chemistry not replenished, developer / fixer levels drift, films degrade over a batch.
Root cause: Worn replenishment-pump tubing or a pump fault.
Symptoms: Films come out damp.
Root cause: Dryer heater or blower fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transport / crossover rollers & gears | Periodic replacement | $$ | Moderate |
| Replenishment-pump tubing | Frequent wear item | $ | Fairly easy |
| Developer heater / thermostat | Major service part | $$ | Moderate |
| Recirculation / replenishment pump | Periodic replacement | $$ | Moderate |
| Dryer heater / blower | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio