Acoma Medical · Model PRO-NEXT
Injection Type VCV, PCV and PSV (Optional) ventilation mode are available. Electronic control PEEP Alarm off function Desflurane anesthesia is available. Flow sensor accuracy is improved. Tidal volume setting can be applied from 30 mL. Functional O2 flush button. Battery backup: 30 minutes Patient unit is equipped with O2 sensor and flow sensor. Space saving design. Injection type vaporizer: Accurate control digital vaporizer adjusts anesthesia concentration by 0.1%. Sevoflurane concentration is available from 0 to 8.0%. Isoflurane concentration is available from 0 to 6.0%. Anesthesia agent bottle can be directly attached to the main unit. Easy and safe handling. Digital flowmeters allows fresh gas flow to be adjusted with 10 mL increments. Backlight is equipped. Consumption of O2, N2O, and anesthesia agent Sevoflurane and Isoflurane can be monitored. Standard type Next generation functional anesthesia machine. VCV, PCV and PSV (Optional) ventilation mode are available. Electronic control PEEP Alarm off function Desflurane anesthesia is available. Flow sensor accuracy is improved. Tidal volume setting can be applied from 30 mL. Functional O2 flush button. Battery backup: 30 minutes Patient unit is equipped with O2 sensor and flow sensor. Space saving design. Vaporizer: 2 units
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Delivered agent concentration off, low/high agent alarm.
Root cause: Worn bellows or O-ring; agent contamination.
Symptoms: O2 low-pressure alarm, inability to ventilate.
Root cause: Pipeline disconnect or failed O2 flush valve.
Symptoms: Low minute volume, volume loss, leak alarm.
Root cause: Loose connections, torn bellows, cracked hose.
Symptoms: High scavenger pressure, agent smell in OR.
Root cause: Obstructed scavenger line or failed regulator.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Requires periodic calibration.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Scavenger interface & hose | Periodic replacement | $$ | Fairly easy |
| O2 pressure regulator | Periodic replacement | $$$ | Difficult |
| Vaporizer (agent-specific) Requires periodic calibration. | Major service part | $$$$ | Expert-level |
| Breathing circuit / bellows | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →