Accriva Diagnostics · Model Hemochron
Hemochron® Report Maker - A data management system for documenting, tracking, and managing patient and QC test results efficiently from all Hemochron devices. Configuration Manager - Configuration management for Hemochron Signature instruments. OxyReviewTM - A report generator for capturing Avoximeter® 1000E results and data.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Fault code on the LCD, a shift in clinical or QC values, or a failed detector-sensitivity test.
Root cause: Loss of the fixed-calibration status - contact ITC / Accriva Technical Service; do not attempt field recalibration.
Symptoms: Daily detector check fails, erratic clot times.
Root cause: Magnetic detector or optics in a test well degraded.
Symptoms: Results drift, temperature verification outside 37 C.
Root cause: Well heater or temperature-control fault.
Symptoms: Low-battery early, will not run a full shift.
Root cause: Aged battery cells or a charger fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Test-well heater / temperature sensor | Major service part | $$ | Moderate |
| Clot-detection (magnetic) detector assembly | Major service part | $$$ | Difficult |
| HEMOCHRON test tubes / QC | Frequent wear item | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio