ABBOTT POINT OF CARE INC. · Model 03P75-82
I-Stat 1 Wireless Analyzer.
Intended use: A portable, electrically-powered instrument intended to be used by health professionals at the point-of-care for the qualitative and/or quantitative in vitro detection of multiple clinical chemistry analytes in a clinical specimen, using one or more technology(ies) [e.g., microfluidics, electrometry, spectrophotometry, colorimetry, fluorimetry, radiometry, chemiluminescence]. Analytes detected may include electrolytes, specific proteins, clinical chemistry substrates (e.g., lipid profile), liver and/or cardiac function test analytes, and/or enzymes.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: "*** " or a numbered cartridge / QC code, result suppressed, run aborts.
Root cause: Cartridge out of date or stored wrong, air in the sample, an under- or over-filled cartridge, or a cartridge sensor fault.
Symptoms: Cartridge not detected, motor / positioning error, will not draw sample.
Root cause: Contaminated or worn internal conductor pins / cartridge port, or a mechanical drive fault.
Symptoms: Low-battery lockout, will not power on, short life on disposable or rechargeable batteries.
Root cause: Depleted or aged batteries, or a battery-contact fault.
Symptoms: Temperature out-of-range code, analyzer will not run cartridges.
Root cause: Ambient outside the operating range, or the internal thermal probe / heater fault.
Symptoms: Results not transmitting to the data manager, will not sync via the downloader or Wi-Fi.
Root cause: Downloader / cradle, network configuration, or the analyzer radio.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Use to verify the analyzer's cartridge-reading electronics.
Single-use; refrigerate and use within the room-temperature window.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Batteries (disposable lithium or rechargeable pack) | Frequent wear item | $ | Easy |
| Electronic Simulator (QC accessory) Use to verify the analyzer's cartridge-reading electronics. | Major service part | $$ | Easy |
| Downloader / recharger cradle | Major service part | $$ | Easy |
| i-STAT cartridges Single-use; refrigerate and use within the room-temperature window. | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.