Mindray · Model BS-240
Fully automated biochemistry analyzer for small-to-medium labs with 240 tests/hour throughput.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Short sample / carryover flags
Root cause: Clotted samples or dried reagent crust
Symptoms: Rising absorbance QC drift, photometer checks fail
Root cause: Lamp output decline over ~2000 hours
Symptoms: Carryover increase, bubbles in cuvettes
Root cause: Tubing loses elasticity
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
PM item every 6 months.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Halogen lamp 12V 100W | Frequent wear item | $80–$150 | Fairly easy |
| Sample probe needle | - | $60–$120 | Moderate |
| Peristaltic tubing set PM item every 6 months. | Frequent wear item | $40–$90 | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.
An issue has been identified with the BS-200 where clinical chemistry results may be inaccurate.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.