Abbott · Model i-STAT 1
Transmits results wirelessly to the electronic medical records (EMR) Provides access to real-time, lab-quality results Offers the broadest menu of tests in a single bedside platform Integrates with users existing i-STAT System data manager and is compatible with 802.11b/802.11g wireless networks.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Results do not fit the patient's clinical picture; repeat on a fresh cartridge disagrees.
Root cause: Interfering substance in the sample, a bad cartridge lot, improper fill, or sample handling delay.
Symptoms: Quality-check code displayed, test aborts, "cartridge error" messages.
Root cause: Contaminated or worn internal contact pins, bent pins in the cartridge port, or a cold/expired cartridge.
Symptoms: Warning at power-on, operator locked out of the Test Menu, date/time not held.
Root cause: Low or depleted battery, real-time-clock coin cell failure, or a condition flagged by the analyzer self-check.
Symptoms: Results not reaching the data manager / EMR, repeated send retries.
Root cause: 802.11b/g Wi-Fi credential or coverage problem, or a data-manager connectivity fault.
Symptoms: Dim or unresponsive screen, missed key presses, barcode will not scan.
Root cause: Aged display, worn keypad membrane, or a dirty / failed scanner window.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Not a repair part; the tool used to verify the reader.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable battery pack | Periodic replacement | $$ | Easy |
| Real-time-clock backup cell | Periodic replacement | $ | Fairly easy |
| Cartridge-port contact pin block | Periodic replacement | $$ | Moderate |
| Electronic Simulator (QC check device) Not a repair part; the tool used to verify the reader. | Major service part | $$$ | Easy |
| Display / keypad assembly | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
From BioLab, for work done with this kind of equipment.