Electrical Safety Analyzer
Walks through the Class I/II, Type B/BF/CF leakage-current test structure - Normal Condition vs Single Fault Condition - with a measured-vs-limit meter for each test.
Enclosure (chassis) leakage
PASS
90 µA
Class
Class I
Applied part
Type BF
Condition
Normal
Insulation
Good
90 uA is within the Normal Condition limit of 500 uA for this test and applied-part type.
Appliance class
Simulated insulation condition
Enclosure (chassis) leakage90 µA / limit 500 µA
PASS
What a BMET verifies with this
- -Protective earth resistance
- -Enclosure and patient (applied-part) leakage current, normal and single-fault
- -Whether the device class and applied-part type change which limit applies
Reference
- Bench analog
- Rigel 288/62353, Fluke ESA612
- Standards referenced
- IEC 60601-1, IEC 62353, NFPA 99
This is a teaching and verification-practice model built from published signal theory and the standards named above, not a calibrated instrument and not a claim about any specific commercial product. Always follow your own analyzer's documentation and your facility's policy for real compliance or acceptance testing.