Electrical Safety Analyzer

Walks through the Class I/II, Type B/BF/CF leakage-current test structure - Normal Condition vs Single Fault Condition - with a measured-vs-limit meter for each test.

Enclosure (chassis) leakage

PASS

90 µA

Class
Class I
Applied part
Type BF
Condition
Normal
Insulation
Good

90 uA is within the Normal Condition limit of 500 uA for this test and applied-part type.

Appliance class
Simulated insulation condition
Enclosure (chassis) leakage90 µA / limit 500 µA

PASS

What a BMET verifies with this

  • -Protective earth resistance
  • -Enclosure and patient (applied-part) leakage current, normal and single-fault
  • -Whether the device class and applied-part type change which limit applies

Reference

Bench analog
Rigel 288/62353, Fluke ESA612
Standards referenced
IEC 60601-1, IEC 62353, NFPA 99

This is a teaching and verification-practice model built from published signal theory and the standards named above, not a calibrated instrument and not a claim about any specific commercial product. Always follow your own analyzer's documentation and your facility's policy for real compliance or acceptance testing.