NIBP Oscillometric Simulator
Simulates a full oscillometric cycle - inflate, step-deflate, and the amplitude envelope the monitor searches - then shows how systolic and diastolic are derived from MAP by characteristic amplitude ratios.
mmHgPress Start to run a full inflate / step-deflate cycle. Set the source pressures below first.
Cuff
0mmHg
MAP (detected)
--mmHg
SYS (detected)
--mmHg
DIA (detected)
--mmHg
Press Start to run a full inflate / step-deflate cycle. Set the source pressures below first.
What a BMET verifies with this
- -Displayed SYS/MAP/DIA match the source within the monitor's rated accuracy
- -Cycle time and cuff over-pressure protection
- -Behavior on a weak or absent oscillation signal
Reference
- Bench analog
- Dynamic NIBP simulators / DPS pressure sources
- Standards referenced
- IEC 80601-2-30
This is a teaching and verification-practice model built from published signal theory and the standards named above, not a calibrated instrument and not a claim about any specific commercial product. Always follow your own analyzer's documentation and your facility's policy for real compliance or acceptance testing.