NIBP Oscillometric Simulator

Simulates a full oscillometric cycle - inflate, step-deflate, and the amplitude envelope the monitor searches - then shows how systolic and diastolic are derived from MAP by characteristic amplitude ratios.

Cuff
0mmHg
MAP (detected)
--mmHg
SYS (detected)
--mmHg
DIA (detected)
--mmHg

Press Start to run a full inflate / step-deflate cycle. Set the source pressures below first.

What a BMET verifies with this

  • -Displayed SYS/MAP/DIA match the source within the monitor's rated accuracy
  • -Cycle time and cuff over-pressure protection
  • -Behavior on a weak or absent oscillation signal

Reference

Bench analog
Dynamic NIBP simulators / DPS pressure sources
Standards referenced
IEC 80601-2-30

This is a teaching and verification-practice model built from published signal theory and the standards named above, not a calibrated instrument and not a claim about any specific commercial product. Always follow your own analyzer's documentation and your facility's policy for real compliance or acceptance testing.