Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
GLOVE PACK 7.0 (15PK), Medline SKU # DYNJ63673D
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Surgeon'S GlovesDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1725-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
1. ANGIO PACK, Medline SKU # DYNJ57032D;
2. NEURO INTERVENTIONAL RADIOLOGY, Medline SKU # DYNJ64026C;
3. NEURO INERVENTIONAL RADIOLOGY, Medline SKU # VASC1650.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Neurological TrayDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1726-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
1. DELIVERY PACK, Medline SKU # DYNJ42892F;
2. DELIVERY PACK, Medline SKU # DYNJ42892G.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Labor And Delivery KitDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1727-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
1. ROBOTIC PACK, Medline SKU # DYNJ58475F;
2. URO/GEN ROBOTIC PACK, Medline SKU # DYNJ60803J.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Gynecological Laparoscopic KitDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1728-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
TAVR PACK, Medline SKU # DYNJ84472
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Eye TrayDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1729-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
1. SHOULDER ARTHROSCOPY PACK, Medline SKU # DYNJ30880M;
2. EXTREMITY PACK, Medline SKU # DYNJ64295A.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Orthopedic TrayDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1730-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Reason: Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
Root cause: Component change control
Recalling firm: Spacelabs Healthcare, Inc.Type: Detector And Alarm, ArrhythmiaDistribution: Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.Recall #Z-1793-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Feb 17, 2026
See Luer Cap Set, MPC-130, set, administration, intravascular
Reason: Molded Products, Inc. is voluntarily conducting a medical device recall of the MPC-130 See Luer Cap Set based on a confirmed complaint of the threaded sleeve not being engaged and becoming un-attached.
Root cause: Process control
Recalling firm: Molded Products IncType: Set, Administration, IntravascularDistribution: US Nationwide distribution in the states of TN, TX, NV, IL, FL.Recall #Z-1774-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Approach CTO-6 Micro Wire Guide, Reference Part Number CMW-14-190-6G, Order Number G50786
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Wire, Guide, CatheterDistribution: Worldwide distribution.Recall #Z-1775-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Wire, Guide, CatheterDistribution: Worldwide distribution.Recall #Z-1776-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Approach CTO-18 Micro Wire Guide, Reference Part Number CMW-14-190-18G, Order Number G50788; Reference Part Number CMW-14-300-18G, Order Number G50792
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Wire, Guide, CatheterDistribution: Worldwide distribution.Recall #Z-1777-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Catheter, Intravascular, DiagnosticDistribution: Worldwide distribution.Recall #Z-1778-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-HC-G-EU, Order Number G57685;
Reference Part Number C-PTIS-100-HC-G-NA, Order Number G57682;
Reference Part Number C-PTIS-100-UNL-HC-G-EU, Order Number G57687;
Reference Part Number C-PTIS-100-UNS-HC-G-EU, Order Number G57686.
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Tube Tracheostomy And Tube CuffDistribution: Worldwide distribution.Recall #Z-1779-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703;
Reference Part Number C-PTISY-100-UNL-HC-G-NA, Order Number G57706;
Reference Part Number C-PTISY-100-UNS-HC-G-NA, Order Number G57704.
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Tube Tracheostomy And Tube CuffDistribution: Worldwide distribution.Recall #Z-1780-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-FLEX-HC, Order Number G32092;
Reference Part Number C-PTIS-100-HC, Order Number G12115.
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Tube Tracheostomy And Tube CuffDistribution: Worldwide distribution.Recall #Z-1781-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Cook Staged Extubation Set: Reference Part Number C-CAE-14.0-83-EXTU, Order Number G24073
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Tube Tracheostomy And Tube CuffDistribution: Worldwide distribution.Recall #Z-1782-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Cook Spectrum Central Venous Catheter Tray: Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0003, Order Number G56424;
Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0018, Order Number G21053;
Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0042, Order Number G34914;
Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0047, Order Number G44431;
Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0048, Order Number G44432.
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 DaysDistribution: Worldwide distribution.Recall #Z-1783-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Cook Strange Bile Duct Stone Exploration Set: Reference Part Number C-CSRC-7.5D-40-STRANGE, Order Number G09108
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Dislodger, Stone, BiliaryDistribution: Worldwide distribution.Recall #Z-1784-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193;
Reference Part Number C-TCCSB-500-UNI-01, Order Number G47665.
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Needle, Emergency AirwayDistribution: Worldwide distribution.Recall #Z-1785-2026
Open, ClassifiedPosted Apr 6, 2026·Initiated Mar 5, 2026
COOK MEDICAL Multipurpose Drainage Catheter: Reference Part Number ULT8.5-38-25-P-CLMS-8-RH-FS, Order Number G29775
Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Root cause: Incorrect or no expiration date
Recalling firm: Cook IncorporatedType: Stents, Drains And Dilators For The Biliary DuctsDistribution: Worldwide distribution.Recall #Z-1786-2026