Open, ClassifiedPosted Apr 10, 2026·Initiated Aug 13, 2025
iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software
Reason: it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.
Root cause: Software design
Recalling firm: Beta Bionics, Inc.Type: Alternate Controller Enabled Insulin Infusion PumpDistribution: US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, MA, MD, ME, MI, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WY AND WASHINGTON DC.Recall #Z-1809-2026
Open, ClassifiedPosted Apr 10, 2026·Initiated Mar 12, 2026
LinkBio CORE Workstation, a component of the CORE Shoulder System. Product Code: 87-9136.
Reason: The CORE Workstation may display an incorrect "Planning Date " when viewing cases on the "Start Case" menu screen.
Root cause: Under Investigation by firm
Recalling firm: Linkbio Corp.Type: Shoulder Arthroplasty Implantation SystemDistribution: US Nationwide distribution in the state of Alabama, Florida, and Kansas.Recall #Z-1810-2026
Open, ClassifiedPosted Apr 10, 2026·Initiated Mar 10, 2026
Arjo Tenor mobile passive patient lift, Model Numbers KHA1000 US and KHA1010 US
Reason: An internal component of the actuators in a certain production batch may not meet the required mechanical strength. This could result, in the worst case scenario, in a sudden loss of the actuator's ability to hold the load, leading to rapid and uncontrolled downward movement of the lifting arm.
Root cause: Process control
Recalling firm: ARJOHUNTLEIGH POLSKA Sp. z.o.o.Type: Lift, Patient, Non-Ac-PoweredDistribution: US Nationwide distribution.Recall #Z-1811-2026
Open, ClassifiedPosted Apr 9, 2026·Initiated Mar 12, 2026
Stryker Sustainability Solution Color Cuff 24" (Yellow) Quick Connect Dual Port Single Bladder REF 5921-024-235
Stryker Sustainability Solution Color Cuff 18" (Red), Quick Connect Dual Port Single Bladder REF 5921-018-135
Stryker Sustainability Solution Color Cuff 18" (Red) Dual Port Single Bladder REF 5921-018-235.
Reason: Due to potential for tourniquet cuffs to detach from the bladder at the welding connection and result in the cuff not being able to maintain pressure.
Root cause: Nonconforming Material/Component
Recalling firm: Stryker Sustainability SolutionsType: Tourniquet, PneumaticDistribution: U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OHJ, OK, OR, PA SC, SD, TN, TX, VA, VT, and WA.
O.U.S.: N/ARecall #Z-1804-2026
Open, ClassifiedPosted Apr 8, 2026·Initiated Sep 8, 2025
MicroScan Neg Multidrug Resistant MIC 1 REF C32368
UDI-DI code: 15099590720568
Intended Use- For use with MicroScan Dried Gram Negative
MIC/Combo Panels and Dried Gram Negative Breakpoint Combo
Panels. MicroScan panels are designed for use in determining
antimicrobial agent susceptibility and/or identification to the species level
of aerobic and facultatively anaerobic gram-negative bacilli.
Reason: Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.
Root cause: Under Investigation by firm
Recalling firm: Beckman Coulter, Inc.Type: Panels, Test, Susceptibility, AntimicrobialDistribution: International distribution to the country of Poland.Recall #Z-1794-2026
Open, ClassifiedPosted Apr 8, 2026·Initiated Mar 12, 2026
B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in vitro diagnostic test for Placental Growth Factor
Reason: Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.
Root cause: Under Investigation by firm
Recalling firm: Brahms GmbHType: Prognostic Test For Development Or Progression Of PreeclampsiaDistribution: Worldwide distribution. US states: NC, TX, MN, UT, NY, FL, OH, IL, CT, CA, GA.
Other countries: DE, AT, AL, AU, BG, BR, CA, CH, CZ, DK, ET, ES, FR, GB, GR, HK, HU, ID, IL, IT, KW, KZ, LT, MK, ML, MX, NL, NO, PA, PL, PT, RO, SE, SI, TH, TU, TW, UA, ZARecall #Z-1795-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Mar 10, 2026
ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
Reason: During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient.
Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system
settings and imaging conditions, as well as the applicable regulatory requirements.
Root cause: Radiation Control for Health and Safety Act
Recalling firm: Siemens Medical Solutions USA, IncType: Interventional Fluoroscopic X-Ray SystemDistribution: Worldwide - US Nationwide distribution.Recall #Z-1653-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
NAMIC Angiographic Syringe, labeled as follows:
1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027;
2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007;
3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107;
4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007;
5. MTO,SYRINGE,12ML,TR/FG,RA,W/RES, Medline SKU 70087107;
6. SYRINGE,12ML,TR/FR,RA,W/O,RES,-,PKG, Medline SKU 70088007;
7. SYRINGE,10ML,PP/FR,RA,-,PKG, Medline SKU 70095007;
8. SYRINGE,10ML,PP/FG,RA,-,PKG, Medline SKU 70095107;
9. SYRINGE,12ML,W/RES,PP/FR,RA,PG, Medline SKU 70097007;
10. SYRINGE,12ML,PP/FG,RA,W/RES,-,PKG, Medline SKU 70097107;
11. SYRINGE,12ML,PP/FG,RA,W/O,RES, Medline SKU 70098107;
12. SYRINGE,20ML,TR/FR,RA,OEM,PG, Medline SKU 80075027;
13. SYRINGE,10ML,ANGIO,W/RA,-,OEM, Medline SKU 80085007;
14. SYRINGE,12ML,TR/FR,RA,OEM, Medline SKU 80087007;
15. SYRINGE,10ML,PP/FR,RA,-,OEM,-,PG, Medline SKU 80095007;
16. SYRINGE,10ML,PP/FG,RA,OEM,PG, Medline SKU 80095107;
17. SYRINGE,12ML,PP/FG,RA,-,OEM,-,PG, Medline SKU 80097107.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Injector And Syringe, AngiographicDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1712-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
NAMIC Angiographic Control Syringe, labeled as follows:
1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007;
2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007;
3. MTS,SYRINGE,8ML,TR/FR,RA,NO,RES,-,P/G, Medline SKU 70084017;
4. SYRINGE,7,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80083007;
5. SYRINGE,8,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80084007;
6. MTO,SYRINGE,8,ML,TR/FR,RA,NO,RES,OEM,PG, Medline SKU 80084017.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Injector And Syringe, AngiographicDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1713-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Nerve Block TrayDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1714-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
1. CSTM MANIFOLD KIT - TITUS, Medline SKU VASC1061;
2. CSTM MANIFOLD KIT - FRYE, Medline SKU VASC1074;
3. CRYO EP MANIFOLD KIT, Medline SKU VASC1198;
4. MANIFOLD KIT - 500PSI OFF, Medline SKU VASCSHPOFF1;
5. MANIFOLD KIT LOW PRESSURE ON, Medline SKU VASCSLPON1.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary BypassDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1715-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
1. CATH LAB PACK-LF, Medline SKU 00-HCT055P;
2. INTERVENTIONAL CATH LAB PK-LF, Medline SKU DYNJ0376230Q;
3. ANGIOGRAPHY CATH LAB, Medline SKU DYNJ23532C;
4. ANGIO LEFT HEART PACK, Medline SKU DYNJ40632C;
5. ANGIO DRAPE PACK, Medline SKU DYNJ47952B;
6. ANGIOGRAPHIC PACK, Medline SKU DYNJ57516;
7. CATH LAB PACK, Medline SKU DYNJ58243B;
8. CATH LAB PACK, Medline SKU DYNJ65185;
9. CSU CATH MANIFOLD KIT, Medline SKU VASC1170A;
10. MANIFOLD KIT NO TRANSDUCER, Medline SKU VASC1184A.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Angiography/Angioplasty KitDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1716-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
NAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107; 4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007; 5. MTO,SYRINGE,12ML,TR/FG,RA,W/RES, Medline SKU 70087107; 6. SYRINGE,12ML,TR/FR,RA,W/O,RES,-,PKG, Medline SKU 70088007; 7. SYRINGE,10ML,PP/FR,RA,-,PKG, Medline SKU 70095007; 8. SYRINGE,10ML,PP/FG,RA,-,PKG, Medline SKU 70095107; 9. SYRINGE,12ML,W/RES,PP/FR,RA,PG, Medline SKU 70097007; 10. SYRINGE,12ML,PP/FG,RA,W/RES,-,PKG, Medline SKU 70097107; 11. SYRINGE,12ML,PP/FG,RA,W/O,RES, Medline SKU 70098107; 12. SYRINGE,20ML,TR/FR,RA,OEM,PG, Medline SKU 80075027; 13. SYRINGE,10ML,ANGIO,W/RA,-,OEM, Medline SKU 80085007; 14. SYRINGE,12ML,TR/FR,RA,OEM, Medline SKU 80087007; 15. SYRINGE,10ML,PP/FR,RA,-,OEM,-,PG, Medline SKU 80095007; 16. SYRINGE,10ML,PP/FG,RA,OEM,PG, Medline SKU 80095107; 17. SYRINGE,12ML,PP/FG,RA,-,OEM,-,PG, Medline SKU 80097107;
18. CATH SET UP PACK, Medline SKU DYNJ46711O
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Cardiac Catheterization KitDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1717-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
NAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107; 4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007; 5. MTO,SYRINGE,12ML,TR/FG,RA,W/RES, Medline SKU 70087107; 6. SYRINGE,12ML,TR/FR,RA,W/O,RES,-,PKG, Medline SKU 70088007; 7. SYRINGE,10ML,PP/FR,RA,-,PKG, Medline SKU 70095007; 8. SYRINGE,10ML,PP/FG,RA,-,PKG, Medline SKU 70095107; 9. SYRINGE,12ML,W/RES,PP/FR,RA,PG, Medline SKU 70097007; 10. SYRINGE,12ML,PP/FG,RA,W/RES,-,PKG, Medline SKU 70097107; 11. SYRINGE,12ML,PP/FG,RA,W/O,RES, Medline SKU 70098107; 12. SYRINGE,20ML,TR/FR,RA,OEM,PG, Medline SKU 80075027; 13. SYRINGE,10ML,ANGIO,W/RA,-,OEM, Medline SKU 80085007; 14. SYRINGE,12ML,TR/FR,RA,OEM, Medline SKU 80087007; 15. SYRINGE,10ML,PP/FR,RA,-,OEM,-,PG, Medline SKU 80095007; 16. SYRINGE,10ML,PP/FG,RA,OEM,PG, Medline SKU 80095107; 17. SYRINGE,12ML,PP/FG,RA,-,OEM,-,PG, Medline SKU 80097107
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Cardiovascular Procedure KitDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1718-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
1. NON VASCULAR PACK, Medline SKU # DYNJ68037C;
2. VEIN, Medline SKU # DYNJ89845;
3. VEIN, Medline SKU # DYNJ89845A.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Central Venous Catheter TrayDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1719-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
SPECIAL PROCEDURE TRAY, Medline SKU # DYNJ80513B
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Ct Biopsy TrayDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1720-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
1. HEAD AND NECK CDS-LF, Medline SKU # CDS780138K;
2. HEART CATH PACK-LF, Medline SKU # DYNJ36478B;
3. HEART CATH PACK-LF, Medline SKU # DYNJ36478B;
4. HEART CATH PACK-LF, Medline SKU # DYNJ36478B;
5. THYROIDECTOMY, Medline SKU # DYNJ910458;
6. LH 3 PORT W/O WASTEBAG, Medline SKU # VASC1081C;
7. LHK, Medline SKU # VASC1510;
8. LHK, Medline SKU # VASC1510.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Ear, Nose, And Throat Surgical TrayDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1721-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
1. LAPAROSCOPIC HERNIORRHAPHY DSC, Medline SKU # DYNJ63021B;
2. LAP PACK, Medline SKU # DYNJ64297B;
3. MINOR PROCEDURE PACK, Medline SKU # DYNJ82411.
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Laparoscopy KitDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1722-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
MINOR PLASTIC-LF, Medline Kit SKU # DYNJ86596
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Plastic Surgery And Accessories KitDistribution: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).Recall #Z-1723-2026
Open, ClassifiedPosted Apr 7, 2026·Initiated Feb 27, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:
1. MTS RH LFT HRT KIT, Medline SKU # 60010508;
2. MTS 3V W/GUIDE, Medline SKU # 60100701;
3. ANGIOGRAPHY KIT, Medline SKU # 60131209;
4. MTS LHK, Medline SKU # 60160075;
5. MTS LHK, Medline SKU # 60161954;
6. MTO LHK, Medline SKU # 60230148;
7. GENERAL SURGERY TRAY, Medline SKU # 00-500510I;
8. BIOBURDEN TEST KIT - 686, Medline SKU # BIOB686;
9. RADIO ANGIO UNIV HOSP PACK-LF, Medline SKU # DYNJ0376145M;
10. SPECIALS PACK-LF, Medline SKU # DYNJ0392706K;
11. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426O;
12. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426P;
13. CARDIAC CATH PACK, Medline SKU # DYNJ07953N;
14. CATH PACK-LF, Medline SKU # DYNJ0803367J;
15. E.P. TRAY, Medline SKU # DYNJ19343J;
16. CARDIAC CATH PACK, Medline SKU # DYNJ20898T;
17. CATH LAB TRAY, Medline SKU # DYNJ24250G;
18. CATH LAB PACK-LF, Medline SKU # DYNJ38333D;
19. CATH LAB PACK-LF, Medline SKU # DYNJ38333F;
20. EP IMPLANT PACK-LF, Medline SKU # DYNJ38910D;
21. ADULT CATH LAB PACK, Medline SKU # DYNJ40380J;
22. CRANIOTOMY PACK-LF, Medline SKU # DYNJ43403D;
23. LAP CHOLE PACK, Medline SKU # DYNJ48487F;
24. CATH LAB ANGIOGRAPHY PACK, Medline SKU # DYNJ50779G;
25. CATH LAB ANGIOGRAPHY PACK, Medline SKU # DYNJ50779I;
26. HYBRID, Medline SKU # DYNJ57049N;
27. SPECIALS PK, Medline SKU # DYNJ57713C;
28. SPECIALS PK, Medline SKU # DYNJ57713D;
29. SPECIALS PK, Medline SKU # DYNJ57713F;
30. CARDIAC CATH LAB, Medline SKU # DYNJ57748A;
31. TAVR PACK, Medline SKU # DYNJ58771G;
32. CATH PACK GWV, Medline SKU # DYNJ59229F;
33. HEART CATH PACK, Medline SKU # DYNJ61294F;
34. CATH LAB PACK, Medline SKU # DYNJ61353C;
35. DEVICE IMPLANT PACK, Medline SKU # DYNJ64636A;
36. CUSTOM PROCEDURE KIT, Medline SKU # DYNJ65277A;
37. GENERAL VASCULAR PACK RFID-V2, Medline SKU # DYNJ66442F;
38. CARDIAC CATH PACK, Medline SKU # DYNJ69845C;
39. OMF MINOR PACK, Medline SKU # DYNJ83991C;
40. HEART CATH PACK, Medline SKU # DYNJ84133B;
41. SPECIAL PROCEDURES, Medline SKU # DYNJ85425B;
42. PRIMARY PACK, Medline SKU # DYNJT3093;
43. CARDIAC CATH PACK, Medline SKU # DYNJT3522A;
44. CARDIAC CATH PACK, Medline SKU # DYNJT3522B;
45. TOTAL KNEE PACK, Medline SKU # DYNJT4069;
46. CARDIAC CATH PACK, Medline SKU # DYNJT5685;
47. CATH PACK, Medline SKU # DYNJT6544;
48. PACK CARDIAC CATH, Medline SKU # DYNJT7591;
49. OBL CATH VEIN PACK, Medline SKU # DYNJT7863;
50. TUMESCENT SYRINGE KIT, Medline SKU # DYNJTUMSYR;
51. CSTM LH KIT, Medline SKU # VASC1054;
52. LHK, Medline SKU # VASC1130A;
53. MANIFOLD KIT, Medline SKU # VASC1137A;
54. LHK, Medline SKU # VASC1299A;
55. LHK, Medline SKU # VASC1399B;
56. LHK, Medline SKU # VASC1403;
57. IR 2 PORT, Medline SKU # VASC1404;
58. LHK, Medline SKU # VASC1405;
59. LHK, Medline SKU # VASC1455A;
60. TWO PORT MANIFOLD KIT, Medline SKU # VASC1496;
61. RHK, US, NAMIC MANIFOLD, Medline SKU # VASC1502;
62. LHK, Medline SKU # VASC1555;
63. 4 VALVE,PERC,MANIFOLD,KIT, Medline SKU # VASC1609;
64. LHK, Medline SKU # VASC1668;
65. LHK, Medline SKU # VASC1694;
66. LHK, Medline SKU # VASC1805;
67. TVR, Medline SKU # VASC1816.