Open, ClassifiedPosted Sep 5, 2024·Initiated Jul 17, 2024
Power Express, REF B90918
Reason: The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
Root cause: Nonconforming Material/Component
Recalling firm: Beckman Coulter Biomedical GmbHType: Radioassay, Vitamin B12Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Bahrain, Canada, China, Colombia, Egypt, France, Germany, Italy, Jordan, Korea, Kuwait, Lebanon, Malaysia, Panama, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, United Arab Amirates, United Kingdom, Vietnam.Recall #Z-2978-2024
Open, ClassifiedPosted Sep 5, 2024·Initiated Jul 17, 2024
PE Centrifuge, REF: B36365, a component of the Power Express
Reason: The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
Root cause: Nonconforming Material/Component
Recalling firm: Beckman Coulter Biomedical GmbHType: Radioassay, Vitamin B12Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Bahrain, Canada, China, Colombia, Egypt, France, Germany, Italy, Jordan, Korea, Kuwait, Lebanon, Malaysia, Panama, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, United Arab Amirates, United Kingdom, Vietnam.Recall #Z-2979-2024
Open, ClassifiedPosted Sep 5, 2024·Initiated Jul 17, 2024
PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.
Reason: The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
Root cause: Nonconforming Material/Component
Recalling firm: Beckman Coulter Biomedical GmbHType: Radioassay, Vitamin B12Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Bahrain, Canada, China, Colombia, Egypt, France, Germany, Italy, Jordan, Korea, Kuwait, Lebanon, Malaysia, Panama, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, United Arab Amirates, United Kingdom, Vietnam.Recall #Z-2980-2024
Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 1, 2024
LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Y-Site (Qty 25), Product Code: SET-0013-25. for intravenous infusion.
Reason: A defect may cause an uncontrolled medication flow condition which may result in a dose greater than prescribed for the patient (i.e., medication overdose).
Root cause: Process control
Recalling firm: Fresenius Kabi USA, LLCType: Set, Administration, IntravascularDistribution: US Distribution to states of: CO, ID, NJ, UT.Recall #Z-2721-2024
Open, ClassifiedPosted Sep 4, 2024·Initiated Jul 22, 2024
Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system.
Part Number 115-066755-00
Reason: V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the
patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe
cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
Root cause: Nonconforming Material/Component
Recalling firm: Mindray DS USA, Inc. dba Mindray North AmericaType: Gas-Machine, AnesthesiaDistribution: US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: CanadaRecall #Z-2949-2024
Open, ClassifiedPosted Sep 4, 2024·Initiated Jul 22, 2024
Sevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system.
Part Number: 115-066756-00
Reason: V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the
patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe
cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
Root cause: Nonconforming Material/Component
Recalling firm: Mindray DS USA, Inc. dba Mindray North AmericaType: Gas-Machine, AnesthesiaDistribution: US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: CanadaRecall #Z-2950-2024
Open, ClassifiedPosted Sep 4, 2024·Initiated Jul 22, 2024
Isoflurane E_Vap.(Safety Filling adapter): used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system.
Part Number: Part Number 115-066758-00
Reason: V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the
patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe
cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
Root cause: Nonconforming Material/Component
Recalling firm: Mindray DS USA, Inc. dba Mindray North AmericaType: Gas-Machine, AnesthesiaDistribution: US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: CanadaRecall #Z-2951-2024
Open, ClassifiedPosted Sep 4, 2024·Initiated Jul 22, 2024
Desflurane E_Vap.(Saf-T-Fill) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system.
Part Number: Part Number115-066759-00
Reason: V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the
patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe
cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
Root cause: Nonconforming Material/Component
Recalling firm: Mindray DS USA, Inc. dba Mindray North AmericaType: Gas-Machine, AnesthesiaDistribution: US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: CanadaRecall #Z-2952-2024
Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 6, 2024
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model Numbers 722079 and 722224
Reason: The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.
Root cause: Under Investigation by firm
Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Type: Interventional Fluoroscopic X-Ray SystemDistribution: Worldwide - US Nationwide distribution.Recall #Z-2953-2024
Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Impella CP Smart Assist Set, EU; Product Code: 0048-0014;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2954-2024 Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Impella CP Smart Assist Set, JP; Product Code: 0048-0024-JP;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2955-2024 Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Impella CP Pump Set, Japan; Product Code: 0048-0034-JP;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2956-2024 Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Impella CP Pump 371 Set, Clinical Trial; Product Code: 0048-0040;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2957-2024 Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Impella CP Smart Assist Set, Canada; Product Code: 0048-0044;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2958-2024 Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Pump 371 14F LT CMR Set; Product Code: 0048-0045;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2959-2024 Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Impella CP Smart Assist Set APAC; Product Code: 0048-0047;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2960-2024 Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Impella 5.5 with SmartAssist Set, EU; Product Code: 0550-0002;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2961-2024 Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Impella 5.5 with SmartAssist Set, CA; Product Code: 0550-0004;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2962-2024 Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Impella 5.5 with SmartAssist Set, US; Product Code: 0550-0008;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2963-2024 Open, ClassifiedPosted Sep 4, 2024·Initiated Aug 5, 2024
Pump 381 Pump Set (US); Product Code: 1000080;
Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Root cause: Under Investigation by firm
Recalling firm: Abiomed, Inc.Type: Temporary Non-Roller Type Left Heart Support Blood PumpDistribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.Recall #Z-2964-2024