CompletedPosted Oct 31, 2024·Initiated Jun 6, 2023
GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers:
1) 2082844-002-01040417;
2) 2082844-002-01044956;
3) 2082844-002-01058424;
4) 2082844-002-01069017;
5) 2082844-002-01089666;
6) 2082844-002-01090474;
7) 2082844-002-01091269;
8) 2082844-002-01093489;
9) 2082844-002-01113262;
10) 2082844-002-01113711;
11) 2082844-002-01123085;
12) 2082844-002-01123918;
13) 2082844-002-01128966;
14) 2082844-002-01133042;
15) 2082844-002-01146447;
16) 2082844-002-100205;
17) 2082844-002-100749;
18) 2082844-002-100752;
19) 2082844-002-100995;
20) 2082844-002-101425;
21) 2082844-002-523548;
22) 2082844-002-529928;
23) 2082844-002-594800;
24) 2082844-002-643544;
25) 2082844-002-925841;
26) 2082844-002-965540;
27) 2082844-002-982916;
28) 2082844-002-983420;
29) 2082844-002-993905;
30) 2082844-002-995018;
31) 2082844-002-995799;
32) 2082844-002-998943; Radiant infant warmer
Reason: Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.
Root cause: Component design/selection
Recalling firm: DATEX--OHMEDA, INC.Type: Incubator, NeonatalDistribution: worldwideRecall #Z-0141-2025
Open, ClassifiedPosted Oct 31, 2024·Initiated Jul 16, 2024
CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-2111-0200-08, 21-2111-0200-12, 21-2111-0200-14, 21-2111-0200-17, 21-2111-0200-232, 21-2111-0200-50, 21-2111-0300-00, 21-2111-0300-01, 21-2111-0300-02, 21-2111-0300-03, 21-2111-0300-06, 21-2111-0300-07, 21-2111-0300-08, 21-2111-0300-09, 21-2111-0300-12, 21-2111-0300-14, 21-2111-0300-17, 21-2111-0300-231, 21-2111-0300-232, 21-2111-0300-50, 21-2111-0400-01, 21-2111-0400-51, 21-2111-0401-02L, 21-2111-0401-03L, 21-2111-0401-06L, 21-2111-0401-07, 21-2111-0401-07L, 21-2111-0401-08L, 21-2111-0401-12L, 21-2111-0401-14L, 21-2111-0401-17L, 21-2111-0401-249, 21-2111-0401-51, 21-2111-0401-51L, 21-2111-0401-78, 21-2111-0402-02, 21-2111-0402-02L, 21-2111-0402-03L, 21-2111-0402-06L, 21-2111-0402-07L, 21-2111-0402-08, 21-2111-0402-08L, 21-2111-0402-09_LOAN, 21-2111-0402-09-FUK, 21-2111-0402-09JP, 21-2111-0402-09-KYO, 21-2111-0402-09-NAG, 21-2111-0402-09-OKA, 21-2111-0402-09-OSA, 21-2111-0402-09-SAP, 21-2111-0402-09-SEN, 21-2111-0402-09-TOKE, 21-2111-0402-09-TOKW, 21-2111-0402-09-YOK, 21-2111-0402-12L, 21-2111-0402-14L, 21-2111-0402-17L, 21-2111-0402-249, 21-2111-0402-51, 21-2111-0402-51L, 21-2111-0402-78, 21-2112-0100-50, 21-2112-0100-51, 21-2112-0200-02, 21-2112-0200-06, 21-2112-0300-00, 21-2112-0300-01, 21-2112-0300-02, 21-2112-0300-03, 21-2112-0300-06, 21-2112-0300-231, 21-2112-0300-50, 21-2112-0400-01, 21-2112-0400-51, 21-2112-0401-02L, 21-2112-0401-03L, 21-2112-0401-07L, 21-2112-0401-08L, 21-2112-0401-17L, 21-2112-0401-51, 21-2112-0401-51L, 21-2112-0401-78, 21-2112-0402-02, 21-2112-0402-02L, 21-2112-0402-03L, 21-2112-0402-06L, 21-2112-0402-07L, 21-2112-0402-08L, 21-2112-0402-09JP, 21-2112-0402-09-SAI, 21-2112-0402-09-SEN, 21-2112-0402-09-TOKE, 21-2112-0402-17L, 21-2112-0402-249, 21-2112-0402-51, 21-2112-0402-51L, 21-2120-0100-02, 21-2120-0100-03, 21-2120-0100-06, 21-2120-0100-232, 21-2120-0100-233, 21-2120-0100-50, 21-2120-0100-51, 21-2120-0100-95, 21-2120-0102-02, 21-2120-0102-03, 21-2120-0102-06, 21-2120-0102-07, 21-2120-0102-08, 21-2120-0102-12, 21-2120-0102-13, 21-2120-0102-14, 21-2120-0102-15, 21-2120-0102-17, 21-2120-0102-247, 21-2120-0102-249, 21-2120-0102-51, 21-2120-0102-78, 21-2120-0102-92, 21-2120-0102-98, 21-2120-0103-01, 21-2120-0103-02, 21-2120-0103-03, 21-2120-0103-06, 21-2120-0103-07, 21-2120-0103-08, 21-2120-0103-14, 21-2120-0103-15, 21-2120-0103-17, 21-2120-0103-249, 21-2120-0103-51, 21-2120-0103-78, 21-2120-0104-01, 21-2120-0105-00, 21-2120-0105-01, 21-2120-0105-02L, 21-2120-0105-03L, 21-2120-0105-06L, 21-2120-0105-07L, 21-2120-0105-08L, 21-2120-0105-12L, 21-2120-0105-13L, 21-2120-0105-14L, 21-2120-0105-15L, 21-2120-0105-17L, 21-2120-0105-231, 21-2120-0105-50, 21-2120-0105-50L, 21-2125-0104-01, 21-2125-0105-07L, 21-2125-0105-08L, 21-2125-0105-50L, 21-2127-0104-01, 21-2127-0105-01, 21-2127-0105-02L, 21-2127-0105-03L, 21-2127-0105-06L, 21-2127-0105-08L, 21-2127-0105-50L, 22-4095-0203-51, 22-4095-0204-25, 22-4095-0205-25.
CompletedPosted Oct 31, 2024·Initiated Sep 27, 2024
AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.
Reason: Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention.
Root cause: Process change control
Recalling firm: Accriva Diagnostics, Inc.Type: OximeterDistribution: US: AZRecall #Z-0279-2025
Open, ClassifiedPosted Oct 31, 2024·Initiated Sep 23, 2024
KLS Martin Battery Pack, REF: KLS BP2 001, and KLS BP2 040 (bulk case of 40), used in conjunction with the KLS Martin MaxDriver 2.0 surgical driver, REF: KLS-SD-2000
Reason: The packaging of a sterile battery, intended to power a surgical screwdriver, may contain Tyvek shavings inside the sterile package.
Root cause: Component design/selection
Recalling firm: Pro-Dex IncType: Battery, Replacement, RechargeableDistribution: US distribution to FL only.Recall #Z-0280-2025
Open, ClassifiedPosted Oct 31, 2024·Initiated Oct 2, 2024
smith&nephew GENESIS II NONPOROUS POSTERIOR STABILIZED FEMORAL COMPONENT, SIZE 5 LEFT, REF 71421015; Uncoated knee femur prosthesis, metallic
Reason: Complaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 5N RIGHT implant instead of a GENESIS II POSTERIOR STABILIZED FEMORAL SIZE 5 LEFT as described on the product label.
Root cause: Packaging change control
Recalling firm: Smith & Nephew IncType: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/PolymerDistribution: US Distribution to states of: MS, TN, NC, CA, GA, DE, AL, and OUS distribution to country of: CanadaRecall #Z-0281-2025
Open, ClassifiedPosted Oct 31, 2024·Initiated Oct 2, 2024
smith&nephew LEGION OXINIUM POSTERIOR STABILIZED NONPOROUS FEMORAL COMPONENT, SIZE 5N, RIGHT, NARROW, REF 71421275; Uncoated knee femur prosthesis, metallic
Reason: Complaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 5N RIGHT implant instead of a GENESIS II POSTERIOR STABILIZED FEMORAL SIZE 5 LEFT as described on the product label.
Root cause: Packaging change control
Recalling firm: Smith & Nephew IncType: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/PolymerDistribution: US Distribution to states of: MS, TN, NC, CA, GA, DE, AL, and OUS distribution to country of: CanadaRecall #Z-0282-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 16, 2024
Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy
Reason: As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.
Root cause: Process control
Recalling firm: Percussionaire CorporationType: Device, Positive Pressure Breathing, IntermittentDistribution: U.S.: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT, CO, LA, VA, CT, SC, NH, NV, NC, NJ, MI, UT, ID, ME, MN, SD, ID, ND, KS, KY, TN, MD, VT, MA, RI,
O.U.S.: Canada, Switzerland (all products shipped OUS remained in Sentec AG's (distributor, parent company of Percussionaire) control, were quarantined, and did not leave the warehouse).Recall #Z-0208-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Aug 23, 2024
Osom Ultra hCG Combo Test- IVD immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine for the early confirmation of pregnancy.
Model/Catalog Number: 1004
Reason: This recall was initiated due to customer complaints of false negative of hCG test results when running patient urine samples. Use of the above listed product lot may result in false negative results. An end user reported that the OSOM Ultra hCG Combo Test, lot 563L13, produced false negative test results from 2 patient urine samples as confirmed by quantitative hCG and ultrasound. During further root cause testing and investigation, an additional 6 devices were found to produce false negative test results from hCG positive samples.
Root cause: Process change control
Recalling firm: Princeton Biomeditech CorpType: Visual, Pregnancy Hcg, Prescription UseDistribution: US Distribution to MO only.Recall #Z-0259-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 3, 2024
Product Name: Insight Base Flex - 8, Ref 608-3TX; The device consists of a sterile Snap that includes reflective features that facilitate imaging of the (attached) instrument using instrument tracking technology when used with a compatible tracking system. Each TrackX Snap has a unique set of 2 or more reflective markers. By having unique configurations of reflective markers, it is possible to identify which TrackX Snap is in the field of view of the camera(s). As such, it is also possible to have multiple TrackX Snaps in the field at once and to have different TrackX Snaps for different surgical instruments. The device is considered to be a sterile accessory.
Reason: The patient label stickers may have the wrong expiration date recorded. The contract manufacturer entered the manufacturing date (2024-04-01) as the expiration date. The actual expiration date of the product is 2027-04-01. This is reflected on the carton label and sterile pouch label correctly.
Root cause: Error in labeling
Recalling firm: Trackx Technology IncType: Interventional Fluoroscopic X-Ray SystemDistribution: NC, VA, FL, PA, NYRecall #Z-0260-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 16, 2024
Ambu VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number 412391000), 41 Fr (Catalog Number 412411000)
Reason: Ambu has received complaints on Ambu VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.
Root cause: Under Investigation by firm
Recalling firm: Ambu Inc.Type: Tube, Tracheal/Bronchial, Differential Ventilation (W/Wo Connector)Distribution: US Nationwide. Canada.Recall #Z-0261-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 26, 2024
Duraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US
Lot CT007694
Reason: Out of specification endotoxin levels.
Root cause: Nonconforming Material/Component
Recalling firm: Integra LifeSciences Corp.Type: Dura SubstituteDistribution: ChinaRecall #Z-0262-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 27, 2024
1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P
2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF 7 5132
The Westmed Flex Extenders are non invasive, single use/disposable, large bore, corrugated plastic tubes that provide a flexible, low resistance lightweight connection between breathing circuits and patient interface (tracheal tube, tracheostomy tube, or
mask).
Reason: Due to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may leak or disconnect and may cause life threating conditions or death.
Root cause: Under Investigation by firm
Recalling firm: Westmed, Inc.Type: Connector, Airway (Extension)Distribution: U.S.: AK, CA, CO, DC, FL, GA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, VA, VT, WA, and WV
O.U.S.: N/ARecall #Z-0263-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 4, 2024
Olympus Electrosurgical Generator, Model Number ESG-410 (SEP) Version WA91327U
Reason: There was an increase in complaints indicating an occurrence of system error E1226 "Could not write to Audio Buffer" during procedures using the ESG-410 (SEP).
Root cause: Software design
Recalling firm: Olympus Corporation of the AmericasType: Electrosurgical, Cutting & Coagulation & AccessoriesDistribution: US Nationwide.Recall #Z-0264-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 5, 2024
Access Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, and A38168.
The Access Thyroglobulin Antibody II (TgAb) assay is used for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems.
Reason: IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results.
Root cause: Error in labeling
Recalling firm: Beckman Coulter, Inc.Type: Immunochemical, Thyroglobulin AutoantibodyDistribution: Distribution was made nationwide to the U.S., including Puerto Rico. There was government distribution but no military distribution. Foreign distribution was made to Mexico.Recall #Z-0265-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 20, 2024
remel BactiDrop Acridine Orange, REF 21502, 50 per package.
Reason: A color change of the solution within the dropper can occur from orange-yellow to light yellow resulting in out-of-specification fluorescent stain results.
Root cause: Under Investigation by firm
Recalling firm: Remel, IncType: Fluorometric, CalciumDistribution: Distribution was made to IL, IN, MN, NE, SD, TX, VA, and WA. There was government distribution but no military distribution.
Foreign distribution was made to the United Kingdom.Recall #Z-0266-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 18, 2024
Nihon Kohden Adult/Pediatric Forehead Disposable SpO2 Sensor, 0.9m Model 809030006
Disp. SpO2 Forehead Probe, Adult, 10/Box
Reason: Due to oximeters not having FDA market approval or clearance to distribute in the U.S.
Root cause: No Marketing Application
Recalling firm: Nihon Kohden America IncType: OximeterDistribution: U.S.: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, KS, MA, MD, MI, MO, MT, NC, ND, NE, NH, OK, PA, SC, TX, VT, WA, and WI.
O.U.S.: N/ARecall #Z-0267-2025 Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 18, 2024
Nihon Kohden Adult Ear Clip SpO2 Sensor, 1.5 meter Model 809030007
Reusable SpO2 Ear Clip, each
Reason: Due to oximeters not having FDA market approval or clearance to distribute in the U.S.
Root cause: No Marketing Application
Recalling firm: Nihon Kohden America IncType: OximeterDistribution: U.S.: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, KS, MA, MD, MI, MO, MT, NC, ND, NE, NH, OK, PA, SC, TX, VT, WA, and WI.
O.U.S.: N/ARecall #Z-0268-2025 Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 26, 2024
AMBU / KING LTSD, SIZE 0,1 W/SUCTION PORT, STERILE- Inserted into a patient s pharynx through the mouth to provide patent airway.
Model/Catalog Number: KLTSD420
Reason: Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway.
Root cause: No Marketing Application
Recalling firm: King Systems Corp. dba Ambu, Inc.Type: Airway, Oropharyngeal, AnesthesiologyDistribution: NationwideRecall #Z-0269-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 26, 2024
KING LTSD,SIZE 1,W/SUCTION PORT STERILE-Inserted into a patient s pharynx through the mouth to provide patent airway.
Model/Catalog Number: KLTSD421
Reason: Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway.
Root cause: No Marketing Application
Recalling firm: King Systems Corp. dba Ambu, Inc.Type: Airway, Oropharyngeal, AnesthesiologyDistribution: NationwideRecall #Z-0270-2025
Open, ClassifiedPosted Oct 30, 2024·Initiated Sep 26, 2024
KING LTSD,SIZE 2,W/SUCTION PORT STERILE-Inserted into a patient s pharynx through the mouth to provide patent airway.
Model/Catalog Number: KLTSD422
Reason: Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway.
Root cause: No Marketing Application
Recalling firm: King Systems Corp. dba Ambu, Inc.Type: Airway, Oropharyngeal, AnesthesiologyDistribution: NationwideRecall #Z-0271-2025