Reason: Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
Root cause: Process change control
Reason: Firm received complaints reporting that guides were warped when pulled out of packaging. This issue may result in extension of surgery to find another product, or may lead to adverse local tissue reaction leading to surgical intervention.
Root cause: Process control
Reason: Potential failure of affected units to recharge the battery of the oxygen concentrators.
Root cause: Nonconforming Material/Component
Reason: Potential failure of affected units to recharge the battery of the oxygen concentrators.
Root cause: Nonconforming Material/Component
Reason: Potential failure of affected units to recharge the battery of the oxygen concentrators.
Root cause: Nonconforming Material/Component
Reason: Potential failure of affected units to recharge the battery of the oxygen concentrators.
Root cause: Nonconforming Material/Component
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control
Reason: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root cause: Process control