Open, ClassifiedPosted Dec 4, 2024·Initiated Nov 5, 2024
GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B79632DA:
EU: RT Refresh Package,
Non-EU: Discovery RT, Optima CT580, or
Discovery CT590 RT; X-ray/computed tomography system
Reason: GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
Root cause: Component design/selection
Recalling firm: GE Healthcare (China) Co., Ltd.Type: System, X-Ray, Tomography, ComputedDistribution: Worldwide distribution.Recall #Z-0613-2025
Open, ClassifiedPosted Dec 4, 2024·Initiated Nov 5, 2024
GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80212DA:
EU: Operator Console (for BrightSpeed,
Elite/BrightSpeed Elite Select),
Non-EU: BrightSpeed Elite, BrightSpeed
Elite Select; X-ray/computed tomography system
Reason: GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
Root cause: Component design/selection
Recalling firm: GE Healthcare (China) Co., Ltd.Type: System, X-Ray, Tomography, ComputedDistribution: Worldwide distribution.Recall #Z-0614-2025
Open, ClassifiedPosted Dec 4, 2024·Initiated Nov 5, 2024
GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80522DA:
China Only: Revolution Discovery CT
Reason: GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
Root cause: Component design/selection
Recalling firm: GE Healthcare (China) Co., Ltd.Type: System, X-Ray, Tomography, ComputedDistribution: Worldwide distribution.Recall #Z-0615-2025
Open, ClassifiedPosted Dec 4, 2024·Initiated Nov 5, 2024
GE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80662DA:
China Only: Discovery CT750HD
Reason: GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
Root cause: Component design/selection
Recalling firm: GE Healthcare (China) Co., Ltd.Type: System, X-Ray, Tomography, ComputedDistribution: Worldwide distribution.Recall #Z-0616-2025
Open, ClassifiedPosted Dec 4, 2024·Initiated Oct 1, 2024
Hollister CalciCare Calcium Alginate Dressing REF 529937R
calcium alginate dressings are indicated for moderate to heavily exuding chronic and acute wounds such as:
pressure ulcers
leg ulcers
cavity wounds
lacerations
post-operative surgical wounds
Reason: Due to defects on primary packaging pouch that compromise the device's ability to maintain a sterile barrier.
Root cause: Packaging process control
Recalling firm: Advanced Medical Solutions Ltd.Type: Dressing, Wound, HydrophilicDistribution: U.S. Nationwide distribution in the states of IL, TX, and VA.Recall #Z-0617-2025
Open, ClassifiedPosted Dec 4, 2024·Initiated Oct 1, 2024
McKesson Calcium Alginate Dressing MFR# 3562
Wound dressing indicated for moderate to heavily exuding chronic and acute wounds, and to control minor bleeding in superficial wounds.
Reason: Due to defects on primary packaging pouch that compromise the device's ability to maintain a sterile barrier.
Root cause: Packaging process control
Recalling firm: Advanced Medical Solutions Ltd.Type: Dressing, Wound, HydrophilicDistribution: U.S. Nationwide distribution in the states of IL, TX, and VA.Recall #Z-0618-2025
Open, ClassifiedPosted Dec 4, 2024·Initiated Oct 1, 2024
MEDLINE MAXORB EXTRA ALGINATE WOUND DRESSING 4X4 SQUARE REF MSC7044EP
MEDLINE MOXORB EXTRA ALBINATE WOUND DRESSING 4X8 RECTANLE REF MSC7048EP
Wound dressing indicated for moderate to heavily exuding chronic and acute wounds, and to control minor bleeding in superficial wounds.
Reason: Due to defects on primary packaging pouch that compromise the device's ability to maintain a sterile barrier.
Root cause: Packaging process control
Recalling firm: Advanced Medical Solutions Ltd.Type: Dressing, Wound, HydrophilicDistribution: U.S. Nationwide distribution in the states of IL, TX, and VA.Recall #Z-0619-2025
Open, ClassifiedPosted Dec 3, 2024·Initiated Aug 8, 2024
Fusion Pro 24, Model 17000
Reason: a. A small number of Fusion Pro 24 units were released with an incorrectly placed Interlock device. As a result, the system can be operated with the front-loading door in an open position.
b. Fails to comply with Class 2 designation, indicated in Part 6 of Laser Product Report. An incorrectly placed Interlock device can allow Class 4 laser radiation fields during operation.
Root cause: Radiation Control for Health and Safety Act
Recalling firm: Epilog Laser Corp.Type: Laser Marker Or Engraver, Industrial Or CommercialDistribution: US Nationwide DistributionRecall #Z-0524-2025
Open, ClassifiedPosted Dec 3, 2024·Initiated Nov 16, 2024
Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.
Reason: Potential for a plan box not updated issue during Interventional procedure with Spectral CT that may lead to a collision with the operator or with the needle placed inside the patient, and an incorrect patient ID software issue that may lead to misdiagnosis.
Root cause: Under Investigation by firm
Recalling firm: Philips North America LlcType: System, X-Ray, Tomography, ComputedDistribution: Worldwide - US Nationwide distribution in the states of AR, AZ, CA, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Belgium, China, Czech Republic, Denmark, Ecuador, France, Germany, Hong Kong, Israel, Italy, Japan, Jordan, Macao, Malaysia, Netherlands, New Zealand, Norway, Palestine, Panama, Poland, Portugal, Singapore, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Vietnam.Recall #Z-0583-2025 Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD EpiCenter Single User Software
Version or Model: 441007
Catalog Number: 441007
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: Calculator/Data Processing Module, For Clinical UseDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD EpiCenter Data Management System
Version or Model: 440887
Catalog Number: 440887
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: Calculator/Data Processing Module, For Clinical UseDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD EpiCenter Data Management System, Multi User Software
Version or Model: 440981
Catalog Number: 440981
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: Calculator/Data Processing Module, For Clinical UseDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD COR PX Instrument
Version or Model: 443988
Catalog Number: 443988
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: Kit, Dna Detection, Human PapillomavirusDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD COR MX Instrument
Version or Model: 443989
Catalog Number: 443989
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: Nucleic Acid Detection System For Non-Viral Microorganism(S) Causing Sexually Transmitted InfectionsDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD COR GX Instrument
Version or Model: 443990
Catalog Number: 443990
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: Nucleic Acid Detection System For Non-Viral Microorganism(S) Causing Sexually Transmitted InfectionsDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD MAX System, BD MAX Instrument
Version or Model: 441916
Catalog Number: 441916
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: C. Difficile Toxin Gene Amplification AssayDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD Synapsys Informatics Solution (Laboratory Solution)
Version or Model: 444150
Catalog Number: 444150
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: Calculator/Data Processing Module, For Clinical UseDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD Phoenix M50 Automated Microbiology System
Version or Model: 443624
Catalog Number: 443624
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: System, Test, Automated, Antimicrobial Susceptibility, Short IncubationDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD BACTEC FX, Instrument Top, Packaged
Version or Model: 441385
Catalog Number: 441385
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: System, Blood CulturingDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Open, ClassifiedPosted Dec 3, 2024·Initiated Oct 31, 2024
BD BACTEC FX40 Instrument
Version or Model: 442296
Catalog Number: 442296
Reason: BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Root cause: Software design
Recalling firm: Becton Dickinson & Co.Type: System, Blood CulturingDistribution: Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.