Open, ClassifiedPosted Feb 26, 2025·Initiated Jan 29, 2025
PENTAX Medical Video Upper GI Scope-Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower gastrointestinal tract
Model: EG29-i20c
Reason: During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.
Root cause: Device Design
Recalling firm: Pentax of America IncType: Endoscope, Accessories, Image Post-Processing For Color EnhancementDistribution: NationwideRecall #Z-1237-2025 Open, ClassifiedPosted Feb 26, 2025·Initiated Jan 29, 2025
Pentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes, monitors and other peripheral devices for endoscopic diagnosis, treatment, and video observation.
Model: EPK-i8020c
Reason: During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.
Root cause: Device Design
Recalling firm: Pentax of America IncType: Endoscope, Accessories, Image Post-Processing For Color EnhancementDistribution: NationwideRecall #Z-1238-2025 Open, ClassifiedPosted Feb 25, 2025·Initiated Feb 3, 2025
LIFEPAK 20e REF 99507-000001
The AED mode is to be used only on patients in cardiopulmonary arrest.
Reason: Incorrect keypad was incorrectly installed onto a defibrillator/monitor.
Root cause: Employee error
Recalling firm: Physio-Control, Inc.Type: Automated External Defibrillators (Non-Wearable)Distribution: U.S.: WV
O.U.S.: N/ARecall #Z-1232-2025 Open, ClassifiedPosted Feb 25, 2025·Initiated Jan 10, 2025
DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.
Reason: Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.
Root cause: Software change control
Recalling firm: BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD.Type: Analyzer, Chemistry (Photometric, Discrete), For Clinical UseDistribution: US: MA, CA, SC, LA, OK, GA, AL, TX, OR, ID, CO, MN, NY, MS, PA, WA, IL, MI, PR.
OUS: Lebanon, Poland, Ghana, Spain, South Africa, Slovakia, Ghana, Croatia, Seychelles, Australia, Republic of KoreaRecall #Z-1233-2025
Open, ClassifiedPosted Feb 24, 2025·Initiated Jan 23, 2025
BD Alaris Pump Module REF 8100
The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration
Reason: Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.
Root cause: Under Investigation by firm
Recalling firm: CareFusion 303, Inc.Type: Pump, InfusionDistribution: U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and Guantanamo Bay, Cuba
O.U.S.: CanadaRecall #Z-1231-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 10, 2025
BD Pyxis: MedFlex, REF: 1119-00; MedBank Tower REF: 1145-00; MedBank MedPass, REF: 139088-01 and associated BD Pyxis user/software/safety guides and quick reference guides.
Reason: Automated dispensing device labeling is being updated to strengthen labeling around manual access to medications to address the risk of potential delays in retrieving contents/ accessing medications that may result from power loss or potential automated dispensing device malfunctions.
Root cause: Labeling design
Recalling firm: CareFusion 303, Inc.Type: Cabinet, Table And Tray, AnesthesiaDistribution: US Nationwide distribution including in the states of AZ, FL, NJ, KY, GA, CA, IL, MD, WI, OH, NC, MS, NY, MO, PA, OR, IN, TX, UT, OK, MI, WA, AL, NV, ME, WV, NE, DC, MA, MT, KS, CO, TN, IA, MN, AR, ID, SC, SD, CT, HI, VA, ND, AK, NM, VT, LA, WY.Recall #Z-1206-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 23, 2025
BD Pyxis MedFlex 2.0 REF 1139-00
Bio-ID Spare Parts REF 353200-01
The BD Pyxis" MedFlex is an Automated Dispensing Cabinet (ADC) intended to securely store and dispense medications and supplies to a qualified and authorized user in a clinical setting.
Reason: Due to fingerprint scanner failing resulting in the scanner heating up.
Root cause: Nonconforming Material/Component
Recalling firm: CareFusion 303, Inc.Type: Cabinet, Table And Tray, AnesthesiaDistribution: Worldwide - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI , IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI, and WY. The countries of Australia, Bahrain, Belgium, Brazil, Canada, Hong Kong, Kuwait, Lebanon, New Zealand, Qatar, Saudi Arabia, Spain, and Thailand.Recall #Z-1207-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Feb 5, 2025
IntelePACS (Image Fusion Module) - InteleViewer
Reason: A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.
Root cause: Software design
Recalling firm: INTELERAD MEDICAL SYSTEMS INCORPORATEDType: System, Image Processing, RadiologicalDistribution: Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, AustraliaRecall #Z-1208-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX - ANATOMIC LAT FIB PLATE 4H LT: Anatomic Lateral Fibula Plate, 4-hole, Left
Intended for Fixation of fractures of the distal tibia
Model/Catalog Number: 770708041
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1209-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX -ANATOMIC LAT FIB PLATE 4H RT: Anatomic Lateral Fibula Plate, 4-hole, Right
Model/Catalog Number: 770708042
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1210-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX -ANATOMIC LAT FIB PLATE 6H LT- Anatomic Lateral Fibula Plate, 6-hole, Left
Intended for Fixation of fractures of the distal tibia
Model/Catalog Number: 770708061
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1211-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 6H RT: Anatomic Lateral Fibula Plate, 6-hole, Right
Model/Catalog Number: 770708062
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1212-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 8H LT: Anatomic Lateral Fibula Plate, 8-hole, Left
Intended to bridge or otherwise stabilize bone fragments to facilitate healing.
Model/Catalog Number: 770708081
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1213-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 8H RT: Anatomic Lateral Fibula Plate, 8-hole, Right
Indicated for use in Fixation of fractures of the distal tibia
Model/Catalog Number: 77070802
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1214-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX- Product Name: ANATOMIC LAT FIB PLATE 10H LT Anatomic Lateral Fibula Plate, 10-Hole, Left.
Intended Fixation of fractures of the distal tibia
Model/Catalog Number: 770708101
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1215-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 10H RT- Anatomic Lateral Fibula Plate, 10-Hole, Right.
Intended forFixation of fractures of the distal tibia
Model/Catalog Number: 770708102
Product Description:
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1216-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX-P: ANATOMIC LAT FIB PLATE 12H LT- Anatomic Lateral Fibula Plate, 12-Hole, Left.
Inended for Fixation of fractures of the distal tibia
Model/Catalog Number: 770708121
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1217-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 12H RT- Anatomic Lateral Fibula Plate, 12-Hole, Right.
Intended for Fixation of fractures of the distal tibia
Model/Catalog Number: 770708122
Product Description:
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1218-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX-MEDIAL TIBIA PLATE 6H LT-Medial Tibia Plate, 6-Hole, Left
Intended for Fixation of fractures of the distal tibia
Model/Catalog Number: 770715061
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1219-2025
Open, ClassifiedPosted Feb 21, 2025·Initiated Jan 21, 2025
A.L.P.S. mvX- MEDIAL TIBIA PLATE 6H RT-Medial Tibia Plate, 6-Hole, Right
Model/Catalog Number: 770715062
Intended for Fixation of fractures of the distal tibia
Reason: Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root cause: Process control
Recalling firm: Tyber MedicalType: Plate, Fixation, BoneDistribution: NationwideRecall #Z-1220-2025