Open, ClassifiedPosted Mar 13, 2025·Initiated Feb 7, 2025
Vista Brite Tip, Judkins Right 4, REF SM7504, JR3 VBT (Special) 110 cm 6F.
Guiding catheter is intended for use for intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vascular systems.
Reason: Catheter manufactured at the incorrect length.
Root cause: Employee error
Recalling firm: Cordis US CorpType: Catheter, PercutaneousDistribution: US: WA
OUS: JapanRecall #Z-1330-2025
Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 13, 2025
Portex Endotracheal Tube, Item Codes 100/105/025 100/105/030 100/105/035 100/111/020 100/111/025 100/111/030 100/111/035 100/111/035 100/126/025 100/126/030 100/126/035 100/127/025 100/127/030 100/127/035 100/141/025 100/141/030 100/141/035
Reason: Affected devices have a smaller diameter than expected and may potentially result in inadequate ventilation to the patient. Patient may experience hypoxia, underdose, or cardiopulmonary collapse.
Root cause: Process change control
Recalling firm: Smiths Medical ASD, Inc.Type: Tube, Tracheal (W/Wo Connector)Distribution: Worldwide - US Nationwide distribution.Recall #Z-1289-2025
Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 13, 2025
Brand Name: ZOLL
Product Name: Powerheart G5 AED
Model/Catalog Number: (1) G5A-00A, (2) G5A-01A, (3) G5A-02A, (4) G5A-03A, (5) G5A-19A, (6) G5A-80A, (7) G5A-80A-TSO.
Software Version: NA
Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process.
Note: Not all configurations described in this document are available in all areas.
Reason: The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
Root cause: Unknown/Undetermined by firm
Recalling firm: ZOLL Medical CorporationType: Automated External Defibrillators (Non-Wearable)Distribution: Domestic: Nationwide Distribution
International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.Recall #Z-1316-2025 Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 13, 2025
Brand Name: ZOLL
Product Name: Powerheart G5 AED
Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO.
Software Version: NA
Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process.
Note: Not all configurations described in this document are available in all areas.
Reason: The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
Root cause: Unknown/Undetermined by firm
Recalling firm: ZOLL Medical CorporationType: Automated External Defibrillators (Non-Wearable)Distribution: Domestic: Nationwide Distribution
International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.Recall #Z-1317-2025 Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 13, 2025
Brand Name: ZOLL
Product Name: Powerheart G5 AED
Model/Catalog Number: (1) G5A-80C, (2) G5A-80C-TSO.
Software Version: NA
Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process.
Note: Not all configurations described in this document are available in all areas.
Reason: The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
Root cause: Unknown/Undetermined by firm
Recalling firm: ZOLL Medical CorporationType: Automated External Defibrillators (Non-Wearable)Distribution: Domestic: Nationwide Distribution
International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.Recall #Z-1318-2025 Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 13, 2025
Brand Name: ZOLL
Product Name: Powerheart G5 AED
Model/Catalog Number: (1) G5A-01C, (2) G5A-02C, (3) G5A-05C, (4) G5A-08C, (5) G5A-10C, (6) G5A-13C, (7) G5A-15C, (8) G5A-22C, (9) G5A-41C, (10) G5A-85C, (11) G5A-90C (12) G5A-80-L;
Software Version: NA
Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process.
Note: Not all configurations described in this document are available in all areas.
Reason: The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
Root cause: Unknown/Undetermined by firm
Recalling firm: ZOLL Medical CorporationType: Automated External Defibrillators (Non-Wearable)Distribution: Domestic: Nationwide Distribution
International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.Recall #Z-1319-2025 Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 13, 2025
Brand Name: ZOLL
Product Name: Powerheart G5 AED
Model/Catalog Number: (1) G5S-01C, (2) G5S-02C, (3) G5S-10C, (4) G5S-11C;
Software Version: NA
Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process.
Note: Not all configurations described in this document are available in all areas.
Reason: The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
Root cause: Unknown/Undetermined by firm
Recalling firm: ZOLL Medical CorporationType: Automated External Defibrillators (Non-Wearable)Distribution: Domestic: Nationwide Distribution
International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.Recall #Z-1320-2025 Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 13, 2025
Brand Name: ZOLL
Product Name: Powerheart G5 AED
Model/Catalog Number: G5S-80C;
Software Version: NA
Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process.
Note: Not all configurations described in this document are available in all areas.
Reason: The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
Root cause: Unknown/Undetermined by firm
Recalling firm: ZOLL Medical CorporationType: Automated External Defibrillators (Non-Wearable)Distribution: Domestic: Nationwide Distribution
International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.Recall #Z-1321-2025 Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 13, 2025
Brand Name: ZOLL
Product Name: Powerheart G5 AED
Model/Catalog Number: (1) G5S-00A-TSO, (2) G5S-01A, (3) G5S-02A, (4) G5S-05A, (5) G5S-08A, (6) G5S-10A, (7) G5S-11A, (8) G5S-80A-TSO, (9) G5S-80-L (10) G5S-90A;
Software Version: NA
Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process.
Note: Not all configurations described in this document are available in all areas.
Reason: The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
Root cause: Unknown/Undetermined by firm
Recalling firm: ZOLL Medical CorporationType: Automated External Defibrillators (Non-Wearable)Distribution: Domestic: Nationwide Distribution
International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.Recall #Z-1322-2025 Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 13, 2025
Brand Name: ZOLL
Product Name: Powerheart G5 AED
Model/Catalog Number: G5S-80A;
Software Version: NA
Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process.
Note: Not all configurations described in this document are available in all areas.
Reason: The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
Root cause: Unknown/Undetermined by firm
Recalling firm: ZOLL Medical CorporationType: Automated External Defibrillators (Non-Wearable)Distribution: Domestic: Nationwide Distribution
International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.Recall #Z-1323-2025 Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 28, 2025
Spectral CT on Rails, Software Version Number 5.1.0, Model number: 728334;
Reason: Multiple problems identified with the software version leading to various scanning and image issues, and unintended device movement.
Root cause: Under Investigation by firm
Recalling firm: Philips Medical Systems Nederland B.V.Type: System, X-Ray, Tomography, ComputedDistribution: Domestic: MNRecall #Z-1324-2025
Open, ClassifiedPosted Mar 12, 2025·Initiated Feb 26, 2025
Brand Name: Da Vinci 5
Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals
Model/Catalog Number: 380730
Software Version: N/A
.
Refer to HHE for additional details.
Component: N/A
Reason: Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
Root cause: Device Design
Recalling firm: Intuitive Surgical, Inc.Type: System, Surgical, Computer Controlled InstrumentDistribution: Worldwide distribution - US Nationwide and the countries of Alaska, Arizona, Arkansas, California, Colorado, Connecticut, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the country of South Korea.Recall #Z-1325-2025
Open, ClassifiedPosted Mar 12, 2025·Initiated Nov 13, 2024
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137
Reason: The DxI 9000 Access Immunoassay
Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results
Root cause: Nonconforming Material/Component
Recalling firm: Beckman Coulter, Inc.Type: Analyzer, Chemistry (Photometric, Discrete), For Clinical UseDistribution: US Nationwide. Global Distribution.Recall #Z-1326-2025
Open, ClassifiedPosted Mar 12, 2025·Initiated Jan 29, 2025
Stryker CinchLock Flex Knotless Anchor with Inserter and Suture, REF CAT02643, Sterile EO, Rx Only MR Safe CE 0197
Reason: Increase in complaints concerning "pullwire" breakage that may result in the broken pullwire remaining in the anchor implant body after deployment.
Root cause: Under Investigation by firm
Recalling firm: Stryker CorporationType: Fastener, Fixation, Nondegradable, Soft TissueDistribution: Worldwide - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the countries of Argentina, Colombia, Netherlands, Spain, Mexico, Australia, United Kingdom, Brazil.Recall #Z-1328-2025
Open, ClassifiedPosted Mar 11, 2025·Initiated Jan 31, 2025
Numerous models of nonsterile hemostatic forceps:
(1) REF BJ500R, Phaneuf Hysterectomy Forceps STR 210MM;
(2) REF BJ501R, Phaneuf Hysterectomy Forceps ANG 205MM;
(3) REF BJ511R, Holzbach Hyst Forceps CVD245MM;
(4) REF BJ520R, Heaney Hyst Forceps Grooved 205MM;
(5) REF BJ521R, Heaney Hyst Forceps Grooved 250MM;
(6) REF BJ522R, Heaney Hyst Forceps Grooved 195MM;
(7) REF BJ524R, Heaney-Rezek Fcpsstr Ser Lgth 210MM;
(8) REF BJ527R, Wiener Hyst Forceps Grooves 240MM;
(9) REF BJ530R, Heaney Hysterectomy Forceps STR 225MM;
(10) REF BJ531R, Heaney Hysterectomy Forceps CVD200MM;
(11) REF BJ532R, Heaney Hysterectomy Forceps CVD215MM;
(12) REF BJ534R, Hysterectomy For.Gwilliamsstr.205MM;
(13) REF BJ550R, Wertheim Parametrium Clamp STR 250MM;
(14) REF BJ551R, Wertheim Parametrium Clamp CVD245MM;
(15) REF BJ552R, Wertheim Parametrium Clamp CVD240MM;
(16) REF BJ553R, Wertheim Parametrium Clamphvy-CVD220MM;
(17) REF MD591, Heany Ballentine Hyst Fcps STR.215MM; and
(18) REF MD592, Heany Ballentine Hyst Fcps CVD.215MM.
Reason: The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
Root cause: Use error
Recalling firm: Aesculap IncType: Clamp, UterineDistribution: Distribution was made nationwide and to PR. There was also government/military distribution.
Foreign distribution was made to Canada, Bermuda, Bahamas, and Qatar.Recall #Z-1308-2025
Open, ClassifiedPosted Mar 11, 2025·Initiated Jan 31, 2025
Numerous models of nonsterile hemostatic forceps:
(1) REF BH100R, Carrel Artery Forceps 85MM;
(2) REF BH104R, Hartmann Mosquito FCSDEL STR 100MM;
(3) REF BH105R, Hartmann Mosquito FCPSDELCVD100MM;
(4) REF BH108R, Micro-Halsted Forceps DEL STR125MM;
(5) REF BH109R, Micro-Halsted Forceps DEL CVM 125mm;
(6) REF BH114R, Hartmann Mosquito FCPSDELSTR1X2100MM;
(7) REF BH115R, Hartmann Mosquito FCPSDELCVD1X2100MM;
(8) REF BH118R, Micro-Halsted FCPS Del STR 1X2125MM;
(9) REF BH119R, Micro-Halsted FCPS Del CVD 1X2125MM;
(10) REF BH120R, Halsted-Mosquito FCPSDELSTR1X2125MM;
(11) REF BH121R, Halsted-Mosquito FCPS DELCVD1X2125MM;
(12) REF BH130R, Providence-Hosp FCPS Del STR140MM;
(13) REF BH131R, Providence-Hosp FCPS Del CVD 140MM;
(14) REF BH134R, Kelly Forceps Del STR 140MM;
(15) REF BH135R, Kelly Forceps Del CVM 140MM;
(16) REF BH139R, Dandy Delicate Forceps CVD140MM;
(17) REF BH140R, Baby Crile Forceps Del STR 140MM;
(18) REF BH141R, Baby Crile FORCEPSDELCVD140MM;
(19) REF BH143R, Dandy Del FCPS LAT-CVD 1X2 140MM;
(20) REF BH144R, Crile Forceps STR 140MM;
(21) REF BH145R, Crile Forceps CVD 140MM;
(22) REF BH150R, Baby Kocher Forceps STR 140MM;
(23) REF BH151R, Kocher Forceps Del CVD 1X2140MM;
(24) REF BH154R, Crile Forceps Del STR 1X2 140MM;
(25) REF BH155R, Crile Forceps Del CVD 1X2 140MM;
(26) REF BH160R, Leriche Delicate Forceps STR 150MM;
(27) REF BH161R, Leriche Delicate Forceps CVD 150MM;
(28) REF BH164R, Rankin FCPS Long Kelly Del STR160MM;
(29) REF BH166R, Crile Forceps STR 160MM;
(30) REF BH170R, Leriche Delicate Forceps STR 1X2150MM;
(31) REF BH171R, Leriche Delicate Forceps CVD1X2150MM;
(32) REF BH176R, Crile Forceps STR 1X2160MM;
(33) REF BH177R, Crile Forceps CVD 1X2 160MM;
(34) REF BH178R, Dixon-Lovelace Artery FCPS.1X2T.STR.160;
(35) REF BH197R, Birkett Delicate Forceps CVD185MM;
(36) REF BH199R, Nissen Delicate Forceps CVP 185MM;
(37) REF BH200R, Adson Delicate Forceps STR185MM;
(38) REF BH201R, Adson Delicate Forceps CVD 185MM;
(39) REF BH202R, Halsted Delicate Forceps STR 185MM;
(40) REF BH203R, Halsted Delicate Forceps CVD 185MM;
(41) REF BH205R, Adson Hemostatic FCPS CVD.215MM;
(42) REF BH206R, Heiss Delicate Forceps STR 200MM;
(43) REF BH207R, Heiss Delicate Forceps CVD 200MM;
(44) REF BH210R, Halsted-Mosquito Forceps DELSTR200MM;
(45) REF BH211R, Halsted-Mosquito FCPS Del CVD 200MM;
(46) REF BH212R, Halsted-Mosquito FCPSDELSTR1X2185MM;
(47) REF BH213R, Halsted-Mosquito FCPSDELCVD1X2185MM;
(48) REF BH214R, Halsted-Mosq.FCPSDELSTR1X2 200MM;
(49) REF BH215R, Halsted-Mosq.FCPSDELCVD 1X2 200MM;
(50) REF BH224R, Sarot Delicate Forceps 240MM;
(51) REF BH225R, Crafoord Delicate Forceps CVD 245MM;
(52) REF BH227R, Artery Forceps Crafoord MODIF.240MM;
(53) REF BH228R, Bengolea Delicate Forceps STR 245MM;
(54) REF BH229R, Bengolea Delicate Forceps CVD 245MM;
(55) REF BH230R, Bridge Delicate Forceps STR 275MM;
(56) REF BH231R, Bridge Delicate Forceps CVD 275MM;
(57) REF BH234R, Sarot Delicate Forceps STR 1X2240MM;
(58) REF BH316R, Pean Artery Forceps STR.165MM;
(59) REF BH332R, Spencer-Wells Forceps STR130MM;
(60) REF BH333R, Spencer-Wells Forceps CVD 130MM;
continued
Open, ClassifiedPosted Mar 11, 2025·Initiated Jan 31, 2025
Numerous models of nonsterile hemostatic forceps:
(1) REF BH198R, Fuchsig Artery Forceps CVD.190MM;
(2) REF BH304R, Pean Artery Forceps Straight 140MM;
(3) REF BH813R, Artery Forceps Curved 250MM;
(4) REF BJ016R, Diss.A.Ligat.Forceps S-Shaped CVD.210MM;
(5) REF BJ020R, Overholt-Geiss Suture FCPS #0CVD195MM;
(6) REF BJ021R, Overholt-Geiss Suture FCPS #1CVD205MM;
(7) REF BJ022R, Overholt-Geiss Suture FCPS #2CVD220MM;
(8) REF BJ023R, Overholt-Geiss Suture FCPS #3 CVD225MM;
(9) REF BJ025R, Overholt-Geiss Suture FCPS #5CVD230MM;
(10) REF BJ026R, Overholt-Geiss Suture FCPS #6 CVD225MM;
(11) REF BJ031R, Overholt-Geiss Suture FCPS#1CVD260MM;
(12) REF BJ034R, Overholt-Geiss Suture FCPS#4CVD270MM;
(13) REF BJ035R, Overholt-Geiss Suture FCPS #5CVD280MM;
(14) REF BJ037R, Forceps CVD Delicate 145MM;
(15) REF BJ038R, Forceps CVD.Delicate 185MM;
(16) REF BJ039R, Forceps CVD.Delicate 235MM;
(17) REF BJ041R, Rumel Suture Forceps SLT-CVD 240MM;
(18) REF BJ042R, Rumel Suture Forceps CVD 240MM;
(19) REF BJ043R, Rumel Suture Forceps MED-CV 240MM;
(20) REF BJ044R, Rumel Suture Forceps HVY-CV 240MM;
(21) REF BJ045R, Rumel Suture Forceps RT-ANG# 5 240MM;
(22) REF BJ047R, Zenker Suture Forceps SLT-CVD 300MM;
(23) REF BJ048R, Zenker Suture Forceps HVY-CVD 290MM;
(24) REF BJ050R, Mixter-O'Shaugnessy Forceps CVD190MM;
(25) REF BJ051R, Wikstroem Suture Forceps RT-ANG 205MM;
(26) REF BJ052R, Finochietto Suture FCPCVDW/HOLE245MM;
(27) REF BJ055R, Mixter Suture Forceps Rt-Ang 230MM;
(28) REF BJ057R, Kantrowitz Thoracic Clamp 200MM;
(29) REF BJ058R, Kantrowitz Thoracic Clamp 245MM;
(30) REF BJ059R, Kantrowitz Thoracic Clamp 275MM;
(31) REF BJ060R, Lahey Forceps Long-Serr 225MM;
(32) REF BJ061R, Mixter Forceps Long-Serr220MM;
(33) REF BJ062R, Mixter Forceps Long-Serr 250MM;
(34) REF BJ068R, Lahey DISS.A.LIGAT.FORCEPS 230MM;
(35) REF BJ070R, Lawrence Suture FCPS DEL CVDD 270MM;
(36) REF BJ073R, Mixter Forceps Long-Serr 290MM;
(37) REF BJ080R, Overholt Forceps # 0 CVD215MM;
(38) REF BJ081R, Overholt Forceps # 1 CVD210MM;
(39) REF BJ082R, Overholt Forceps # 2 CVD 220MM;
(40) REF BJ083R, Overholt Forceps # 3 CVD 225MM;
(41) REF BJ086R, Overholt Forceps # 6 CVD 225MM;
(42) REF BJ090R, Overholt Forceps # 0 CVD 295MM;
(43) REF BJ091R, Overholt Forceps # 1 CVD 295MM;
(44) REF BJ096R, Overholt Forceps # 6 CVD 300MM;
(45) REF BJ100R, Gemini Clamp RT-ANG140MM;
(46) REF BJ102R, Gemini Clamp RT-ANG 180MM;
(47) REF BJ103R, Gemini Clamp RT-ANG 200MM;
(48) REF BJ104R, Gemini Clamp RT-ANG 230MM;
(49) REF BJ106R, Gemini Clamp RT-ANG 280MM;
(50) REF BJ120R, O'Shaugnessy DISS.A.LIGAT.FORCEPS 160MM;
(51) REF BJ121R, O'Shaugnessy DISS.A.LIGAT.FORCEPS 180MM;
(52) REF BJ122R, O'Shaugnessy DISS.A.LIGAT.FORCEPS 200MM;
(53) REF BJ123R, O'Shaugnessy DISS.A.LIGAT.FORCEPS 220MM;
(54) REF BJ130R, Zenker DISS & LIG FCPSSLT-CVD 350MM;
(55) REF BJ131R, Zenker DISS & LIG FCPS CVD 350MM;
continued
Open, ClassifiedPosted Mar 11, 2025·Initiated Jan 31, 2025
Numerous models of nonsterile hemostatic forceps:
(1) REF FB458R, Glover ATR.CLAMP CVD 215MM;
(2) REF FB459R, Glover ATR.CLAMP CVD 245MM;
(3) REF FB461R, Glover ATR.CLAMP 195MM;
(4) REF FB462R, Glover ATR.CLAMP 210MM;
(5) REF FB469R, Leland-Jones Peripheral CLMPSTR 195MM;
(6) REF FB567R, Cooley Pediatric Clamp 30DG30/165MM;
(7) REF FB568R, Cooley Pediatric Clamp 60DG 30/160MM;
(8) REF FB569R, Cooley Pediatric Clamp 90DG30/150MM;
(9) REF FB729R, Cooley Pediatric Clamp 17MM 170MM;
(10) REF FB730R, Cooley Pediatric Clamp 20MM175MM;
(11) REF FB736R, Cooley Pediatric Clamp 74MM 185MM;
(12) REF FB737R, Cooley Pediatric Clamp 86MM 195MM; and
(13) REF FB738R, Cooley Pediatric Clamp 108MM 260MM.
Reason: The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
Root cause: Use error
Recalling firm: Aesculap IncType: ClampDistribution: Distribution was made nationwide and to PR. There was also government/military distribution.
Foreign distribution was made to Canada, Bermuda, Bahamas, and Qatar.Recall #Z-1311-2025
Open, ClassifiedPosted Mar 11, 2025·Initiated Jan 31, 2025
Numerous models of nonsterile hemostatic forceps:
(1) REF BH951R, Negus Tonsil Forceps HVY-CVD 190MM;
(2) REF BH952R, Negus Tonsil Forceps X-HVY-CVD190MM;
(3) REF BH957R, Schnidt Tonsil Forceps SLT-CVD185MM;
(4) REF BH959R, Schnidt Tonsil Forceps CVD 185MM;
(5) REF BH961R, Schnidt Tonsil FCPSOPEN-RINGCVD 185MM;
(6) REF BH963R, Schnidt Tonsil FCPSOPEN-RINGCVD185MM; and
(7) REF BH965R, Wilson Tonsil Forceps 190MM.
Reason: The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
Root cause: Use error
Recalling firm: Aesculap IncType: Forceps, EntDistribution: Distribution was made nationwide and to PR. There was also government/military distribution.
Foreign distribution was made to Canada, Bermuda, Bahamas, and Qatar.Recall #Z-1312-2025
Open, ClassifiedPosted Mar 11, 2025·Initiated Dec 27, 2023
Brand Name: TruScan Body, REF CN38-SB
Product Name: TruScan Body compatible with CONELOG 3.8
Model/Catalog Number: CN38-SB
Software Version: N/A
Reason: Retroactive: Due to defect and quality issues dental scanner may not scan properly/accurately
Root cause: Unknown/Undetermined by firm
Recalling firm: TruAbutment Inc.Type: Accessories, Implant, Dental, EndosseousDistribution: US: CA, CO, FL, GA, IA, IL, MD, MN, MS, OH, OR, SD, UT, VA, WA, WI
OUS: Japan, Great Britain , CanadaRecall #Z-1313-2025