Open, ClassifiedPosted Nov 14, 2025·Initiated Sep 30, 2025
MEDLINE convenience kits labeled as:
1) MAJOR VAGINAL HARPER PACK-LF, REF DYNJ0161768G;
2) ACH MAJOR VAGINAL PACK, REF DYNJ67757B.
Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Labor And Delivery KitDistribution: US Nationwide distribution.Recall #Z-0523-2026
Open, ClassifiedPosted Nov 14, 2025·Initiated Sep 30, 2025
MEDLINE convenience kits labeled as:
1) GYN LAPAROSCOPY CDS #34-RF, REF CDS920099S;
2) GYN LAPAROSCOPY CDS-LF, REF CDS983764J;
3) ROBOTIC, REF CDS984262N;
4) ROBOTIC, REF CDS984262P;
5) ROBOTIC UROLOGY-RF, REF DYNJ38844N;
6) DA VINCI PROSTATE/COLPOPEXY, REF DYNJ44864S;
7) LAP-GYN PACK, REF DYNJ44876M;
8) ROBOT PACK, REF DYNJ45184F;
9) GYN LAP PACK, REF DYNJ52524D;
10) GYN LAP PACK, REF DYNJ64234C;
11) GYN DAVINCI PACK, REF DYNJ65373A;
12) ROBOT PACK, REF DYNJ67190C;
13) ROBOTIC PACK, REF DYNJ86154;
14) GYN LAPAROTOMY, REF DYNJ900296K;
15) ROBOT THOR, REF DYNJ900300K;
16) ROBOT URO, REF DYNJ900301K;
17) SM DAVINCI, REF DYNJ903539G;
18) ROBOTIC PACK-LF, REF DYNJ904250I;
19) ROBOTIC-LF, REF DYNJ907910B;
20) GP-ROBOTIC PACK-LF, REF PHS504647004G.
Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Gynecological Laparoscopic KitDistribution: US Nationwide distribution.Recall #Z-0524-2026
Open, ClassifiedPosted Nov 14, 2025·Initiated Sep 30, 2025
MEDLINE convenience kits labeled as:
CIRCUMCISION PACK, REF DYNJ16826O
Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Circumcision TrayDistribution: US Nationwide distribution.Recall #Z-0525-2026
Open, ClassifiedPosted Nov 14, 2025·Initiated Sep 30, 2025
MEDLINE convenience kits labeled as:
1) C-SECTION OR 30 CDS, REF CDS820075W;
2) C-SECTION, REF CDS984261J;
3) C-SECTION CDS, REF CDS985282F;
4) C-SECTION PACK, REF DYNJ02577D;
5) C-SECTION PACK-LF, REF DYNJ0536924U;
6) C-SECTION PACK-LF, REF DYNJ0551169V;
7) C-SECTION PACK-LF, REF DYNJ0551409P;
8) C-SECTION PACK-LF, REF DYNJ07836R;
9) C-SECTION, REF DYNJ14103C;
10) C-SECTION PACK, REF DYNJ19417R;
11) C-SECTION PACK-LF, REF DYNJ33059B;
12) C-SECTION PACK SC-LF, REF DYNJ38063D;
13) C-SECTION PACK SC-LF, REF DYNJ38700G;
14) C-SECTION PACK-LF, REF DYNJ38845D;
15) C-SECTION PACK-LF, REF DYNJ38845F;
16) C-SECTION PACK, REF DYNJ44679L;
17) C-SECTION DELIVERY PACK-LF, REF DYNJ47808F;
18) C SECTION PACK, REF DYNJ48087I;
19) C-SECTION I, REF DYNJ51354B;
20) C-SECTION SAFETY PACK, REF DYNJ52481B;
21) C-SECTION SMMC, REF DYNJ59457B;
22) C-SECTION PACK, REF DYNJ59825A;
23) PK CUST C-SECTION ST MICH, REF DYNJ60544C;
24) C SECTION PACK, REF DYNJ64377D;
25) C-SECTION PACK, REF DYNJ64505B;
26) C SECTION SHARED, REF DYNJ64775C;
27) LDRP-C SECTION PACK-LF, REF DYNJ64834A;
28) LDRP-C SECTION PACK-LF, REF DYNJ64834B;
29) C-SECTION PACK-LF, REF DYNJ65031D;
30) C SECTION PK, REF DYNJ65403A;
31) C SECTION PACK-LF, REF DYNJ65765A;
32) C-SECTION PACK-LF, REF DYNJ69445;
33) C-SECTION PACK-LF, REF DYNJ69445A;
34) C SECTION DELIVERY PACK, REF DYNJ81604A;
35) C SECTION DELIVERY PACK, REF DYNJ81604B;
36) C SECTION PACK, REF DYNJ82949;
37) C-SECTION ADD ON PACK, REF DYNJ83507;
38) C-SECTION PACK, REF DYNJ83980;
39) C SECTION ACCESS, REF DYNJ85612;
40) ISSAQUAH C-SECTION PACK-LF, REF DYNJ85770;
41) C-SECTION PACK, REF DYNJ86173;
42) C-SECTION, REF DYNJ900266F;
43) C-SECTION, REF DYNJ903865D;
44) C-SECTION, REF DYNJ904612G;
45) C-SECTION, REF DYNJ904612I;
46) C-SECTION, REF DYNJ904653G;
47) C-SECTION, REF DYNJ904697F;
48) CDS C-SECTION CH, REF DYNJ906263D;
49) C SECTION, REF DYNJ908327;
50) C-SECTION -LF, REF DYNJ9932207R;
51) C-SECTION -LF, REF DYNJ9932207S;
52) C-SECTION PACK, REF DYNJV0322A;
53) GP-C SECTION PACK-LF, REF PHS396980007C.
Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Open, ClassifiedPosted Nov 14, 2025·Initiated Sep 30, 2025
MEDLINE convenience kits labeled as:
1) GYN LAPAROTOMY HARPER PACK-LF, REF DYNJ0160684F;
2) MAJOR LITHOTOMY PACK-LF, REF DYNJ0425778O;
3) MINOR VAGINAL #76-RF, REF DYNJ27434R;
4) ABDOMINAL HYST PACK, REF DYNJ44848M;
5) DA VINCI HYSTERECTOMY, REF DYNJ44863Q;
6) VAGINAL HYSTERECTOMY PACK, REF DYNJ44900P;
7) URO GYN, REF DYNJ50047G;
8) ISS VAG HYST PACK, REF DYNJ50315F;
9) PK MAJOR GYN ST MICHAEL C, REF DYNJ60554B;
10) VAGINAL HYSTERECTOMY PACK, REF DYNJ67194B;
11) LAVH, REF DYNJ900298L;
12) LITHOTOMY-LF, REF DYNJ900721C;
13) GP-T A H PACK-LF, REF PHS396954006G;
14) GP-VAG HYSTERECTOMY PACK-LF, REF PHS396975006F.
Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Hysterectomy KitDistribution: US Nationwide distribution.Recall #Z-0527-2026
Open, ClassifiedPosted Nov 14, 2025·Initiated Sep 30, 2025
MEDLINE convenience kits labeled as: GP-LEEP PACK-LF, REF PHS396969004F
Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Obstetrical KitDistribution: US Nationwide distribution.Recall #Z-0528-2026
Open, ClassifiedPosted Nov 14, 2025·Initiated Sep 30, 2025
MEDLINE convenience kits labeled as:
1) MINOR EYE TRAY #81, REF DYNJ27466S;
2) MINOR EYE TRAY #81, REF DYNJ27466T.
Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Root cause: Under Investigation by firm
Recalling firm: Medline Industries, LPType: Eye TrayDistribution: US Nationwide distribution.Recall #Z-0529-2026
Open, ClassifiedPosted Nov 14, 2025·Initiated Sep 30, 2025
MEDLINE convenience kits labeled as:
1) NEURO CRANIOTOMY CDS #36-RF, REF CDS780119V;
2) BASIC BACK CDS, REF CDS780147N;
3) INSTRUMENTED BACK CDS, REF CDS780148R;
4) NEURO-LAMI CDS, REF CDS780208I;
5) TOTAL HIP, REF CDS920027Y;
6) TOTAL KNEE, REF CDS940072Y;
7) TOTAL KNEE REPLACEMENT CDS, REF CDS940823AB;
8) TOTAL HIP CDS, REF CDS940843AA;
9) TOTAL HIP CDS, REF CDS940963Y;
10) SHOULDER CDS, REF CDS941065L;
11) LAMINECTOMY CDS-LF, REF CDS981923X;
12) TOTAL KNEE CDS, REF CDS982823S;
13) LOWER EXTREMITY CDS, REF CDS982943S;
14) LOWER EXTREMITY CDS, REF CDS982943T;
15) TOTAL KNEE, REF CDS982986U;
16) EXTREMITY CDS, REF CDS983031Q;
17) PODIATRY CDS, REF CDS983610L;
18) TOTAL HIP CDS-LF, REF CDS983761N;
19) BASIC ORTHO CDS, REF CDS983769L;
20) UPPER EXTREMITY CDS, REF CDS983907L;
21) LOWER EXTREMITY CDS, REF CDS983909P;
22) GENERAL ORTHO CDS, REF CDS983915M;
23) MAJOR ORTHO, REF CDS984253L;
24) LOWER EXTREMTIY, REF CDS984254M;
25) TOTAL KNEE, REF CDS985566R;
26) ORTHO PACK, REF DYNJ00284M;
27) CRANIOTOMY PACK-LF, REF DYNJ0160354C;
28) CRANIOTOMY PACK-LF, REF DYNJ0161342C;
29) CRANIOTOMY PACK-LF, REF DYNJ0161342D;
30) CRANIOTOMY HARPER PACK-LF, REF DYNJ0162022G;
31) SPINE PACK-LF, REF DYNJ0376635P;
32) GP-TOTAL JOINT -LF, REF DYNJ0396868V;
33) GP-EXTREMITY/SHOULDER PACK-LF, REF DYNJ0396906P;
34) EXTREMITY PACK-LF, REF DYNJ0519273N;
35) TOTAL KNEE PACK-LF, REF DYNJ0530895AO;
36) TOTAL KNEE PACK-LF, REF DYNJ0530895AP;
37) UPPER EXTREMITY PACK-LF, REF DYNJ0532846S;
38) LOWER EXTREMITY PACK-LF, REF DYNJ0532852Q;
39) ACL SSC PACK-LF, REF DYNJ0536892U;
40) TOTAL HIP PACK-LF, REF DYNJ0536903AL;
41) TOTAL HIP PACK-LF, REF DYNJ0536903AM;
42) LOWER EXTREMITY PACK-LF, REF DYNJ0587115R;
43) TOTAL HIP PACK-LF, REF DYNJ0618354V;
44) TOTAL KNEE PACK-LF, REF DYNJ0646272S;
45) SPINE ANT POST ADD A PACK-LF, REF DYNJ0753208J;
46) OPEN SHOULDER SSC PACK, REF DYNJ0840659T;
47) TOTAL JOINT PACK-LF, REF DYNJ0885263F;
48) TOTAL KNEE PACK, REF DYNJ21813J;
49) CRANIO/MAXILOFACIAL #49-RF, REF DYNJ21826R;
50) TOTAL HIP PACK, REF DYNJ27315K;
51) PODIATRY PACK, REF DYNJ33978K;
52) LOWER EXTREMITY, REF DYNJ34420N;
53) LOWER EXTREMITY, REF DYNJ34420O;
54) GENERAL ORTHO, REF DYNJ34428O;
55) ARTHROSCOPY PACK, REF DYNJ35835G;
56) HAND PACK, REF DYNJ37697J;
57) ACH SPINE PACK, REF DYNJ38042K;
58) SHOULDER ARTHROSCOPY PACK, REF DYNJ38224C;
59) LAMINECTOMY PACK, REF DYNJ38227C;
60) EXTREMITY PACK, REF DYNJ38234G;
61) TOTAL JOINT PACK, REF DYNJ40982D;
62) SPINE PACK, REF DYNJ40995B;
63) EXTREMITY PACK, REF DYNJ41416G;
64) TOTAL JOINT PACK, REF DYNJ42793D;
65) CRANIOTOMY PACK-LF, REF DYNJ43208G;
66) HIP I PACK-LF, REF DYNJ43215I;
67) LAMINECTOMY PACK-LF, REF DYNJ43218G;
68) LIMB PACK-LF, REF DYNJ43220G;
69) SPLIT PACK-LF, REF DYNJ43225G;
70) TOTAL KNEE PACK-LF, REF DYNJ43226I;
71) EXTREMITY PACK, REF DYNJ44681J;
72) SHOULDER PACK, REF DYNJ44685L;
73) TOTAL KNEE PACK, REF DYNJ44687L;
74) UPPER EXTREMITY PACK, REF DYNJ44688I;
75) PODIATRY PACK, REF DYNJ44692G;
76) ACL PACK, REF DYNJ44849F;
77) CENTRACARE PLAZA-MINOR ORTHO, REF DYNJ44860I;
78) MAJOR EXTREMITY, REF DYNJ44878D;
79) ORTHO HIP, REF DYNJ44884G;
80) EXTREMITY PACK, REF DYNJ45332G;
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 2, 2025
CEREPAK Detachable Coil System - Name/REF:
UNIFORM 2MM X 2.5 CM/FCX100202,
UNIFORM 2.5MM X 3.3 CM/FCX100253,
UNIFORM 3MM X 5.4 CM/FCX100305,
UNIFORM 3.5MM X 6.6 CM/FCX100356,
UNIFORM 4MM X 7.5 CM/FCX100407,
UNIFORM 4MM X 11.5 CM/FCX100412,
UNIFORM 5MM X 9.7 CM/FCX100509,
UNIFORM 5MM X 17 CM/FCX100517,
UNIFORM 6MM X 11.9 CM/FCX100611,
UNIFORM 6MM X 26 CM/FCX100626,
UNIFORM 7MM X 13.9 CM/FCX100713,
UNIFORM 7MM X 30 CM/FCX100730,
UNIFORM 8MM X 16.1 CM/FCX100816,
UNIFORM 8MM X 29 CM/FCX100829,
UNIFORM 3D 2MM X 2.5 CM/FCX140225,
UNIFORM 3D 3MM X 3 CM/FCX140303,
UNIFORM 3D 3MM X 4 CM/FCX140304,
UNIFORM 3D 3MM X 6 CM/FCX140306,
UNIFORM 3D 4MM X 6 CM/FCX140406,
UNIFORM 3D 4MM X 8 CM/FCX140408,
UNIFORM 3D 4MM X 4.5 CM/FCX140445,
UNIFORM 3D 5MM X 7 CM/FCX140507,
UNIFORM 3D 5MM X 12 CM/FCX140512,
UNIFORM 3D 6MM X 9 CM/FCX140609,
UNIFORM 3D 6MM X 15 CM/FCX140615,
UNIFORM 3D 7MM X 11 CM/FCX140711,
UNIFORM 3D 7MM X 17 CM/FCX140717,
UNIFORM 3D 8MM X 12 CM/FCX140812,
UNIFORM 3D 8MM X 20 CM/FCX140820,
UNIFORM 3D 9MM X 14 CM/FCX140914,
UNIFORM 3D 9MM X 22 CM/FCX140922,
UNIFORM 3D 10MM X 16 CM/FCX141016,
UNIFORM 3D 10MM X 25 CM/FCX141025,
UNIFORM 3D 11MM X 18 CM/FCX141118,
UNIFORM 3D 11MM X 27 CM/FCX141127,
UNIFORM 3D 12MM X 19 CM/FCX141219,
UNIFORM 3D 12MM X 30 CM/FCX141230,
UNIFORM XL 5MM X 10 CM/FCX180510,
UNIFORM XL 6MM X 12.2 CM/FCX180612,
UNIFORM XL 7MM X 14.3 CM/FCX180714,
UNIFORM XL 8MM X 13.6 CM/FCX180813,
UNIFORM XL 8MM X 30 CM/FCX180830,
UNIFORM XL 9MM X 15.3 CM/FCX180915,
UNIFORM XL 9MM X 33 CM/FCX180933,
UNIFORM XL 10MM X 17.1 CM/FCX181017,
UNIFORM XL 10MM X 34 CM/FCX181034,
UNIFORM XL 11MM X 18.9 CM/FCX181118,
UNIFORM XL 11MM X 37 CM/FCX181137,
UNIFORM XL 12MM X 20.7 CM/FCX181220,
UNIFORM XL 12MM X 40 CM/FCX181240,
UNIFORM XL 13MM X 43 CM/FCX181343,
UNIFORM XL 14MM X 47 CM/FCX181447,
UNIFORM XL 15MM X 50 CM/FCX181550,
UNIFORM XL 16MM X 47 CM/FCX181647,
UNIFORM XL 17MM X 50 CM/FCX181750,
UNIFORM XL 18MM X 46 CM/FCX181846,
UNIFORM XL 19MM X 50 CM/FCX181950,
UNIFORM XL 20MM X 50 CM/FCX182050,
FREEFORM MINI 1MM X 1CM/MCR091010,
FREEFORM MINI 1MM X 1.5CM/MCR091015,
FREEFORM MINI 1MM X 2CM/MCR091020,
FREEFORM MINI 1MM X 2.5CM/MCR091025,
FREEFORM MINI 1MM X 3CM/MCR091030,
FREEFORM MINI 1MM X 4CM/MCR091040,
FREEFORM MINI 1.5MM X 2CM/MCR091520,
FREEFORM MINI 1.5MM X 2.5CM/MCR091525,
FREEFORM MINI 1.5MM X 3CM/MCR091530,
FREEFORM MINI 1.5MM X 4CM/MCR091540,
FREEFORM MINI 2MM X 2CM/MCR092020,
FREEFORM MINI 2MM X 3CM/MCR092030,
FREEFORM MINI 2MM X 4CM/MCR092040,
FREEFORM MINI 2MM X 6CM/MCR092060,
FREEFORM MINI 2.5MM X 3.5CM/MCR092535,
FREEFORM MINI 2.5MM X 4.5CM/MCR092545,
FREEFORM MINI 2.5MM X 5.5CM/MCR092555,
FREEFORM MINI 3MM X 4CM/MCR093040,
FREEFORM MINI 3MM X 6CM/MCR093060,
FREEFORM MINI 3MM X 8CM/MCR093080,
FREEFORM 2MM X 2 CM/SCR120202,
FREEFORM 3MM X 6 CM/SCR120306,
FREEFORM 3MM X 8 CM/SCR120308,
FREEFORM 4MM X 7 CM/SCR120407,
FREEFORM 4MM X 10 CM/SCR120410,
FREEFORM 4MM X 12 CM/SCR120412,
FREEFORM 5MM X 10 CM/SCR120510,
FREEFORM 5MM X 15 CM/SCR120515,
FREEFORM 6MM X 10 CM/SCR120610,
FREEFORM 6MM X 15 CM/SCR120615,
FREEFORM 6MM X 20 CM/SCR120620,
FREEFORM 7MM X 15 CM/SCR120715,
FREEFORM 7MM X 21 CM/SCR120721,
FREEFORM 8MM X 15 CM/SCR120815,
FREEFORM 8MM X 24 CM/SCR120824,
FREEFORM 9MM X 25 CM/SCR120925,
FREEFORM 10MM X 30 CM/SCR121030,
FREEFORM 12MM X 30 CM/SCR121230,
FREEFORM 2.5MM X 5 CM/SCR122505,
FREEFORM 2.5MM X 3.5 CM/SCR122535,
FREEFORM 3.5MM X 9 CM/SCR123509,
FREEFORM 3.5MM X 7.5 CM/SCR123575,
HELIFORM SOFT 1.5MM X 4 CM/SHD100154,
HELIFORM SOFT 2MM X 8 CM/SHD100208,
HELIFORM SOFT 2.5MM X 10 CM/SHD100250,
HELIFORM SOFT 2.5MM X 8 CM/SHD100258,
HELIFORM SOFT 3MM X 8 CM/SHD100308,
HELIFORM SOFT 3MM X 10 CM/SHD100310,
HELIFORM SOFT 4MM X 8 CM/SHD100408,
HELIFORM SOFT 4MM X 10 CM/ SHD100410,
HELIFORM SOFT 5MM X 10 CM/ SHD100510,
HELIFORM SOFT 5MM X 15 CM/ SHD100515,
HELIFORM SOFT 6MM X 12 CM/ SHD100612,
HELIFORM SOFT 6MM X 16 CM/ SHD100616,
HELIFORM SOFT 7MM X 16 CM/ SHD100716,
HELIFORM SOFT 7MM X 20 CM/ SHD100720,
HELIFORM SOFT 8MM
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 10, 2025
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits;
Product description/Product numbers:
ADULT 120 EXPANDABLE W/MASK
DYNJAA10591;
ANES CIRC 120IN EXP-LF
DYNJAA6555G;
ANES CIRCUIT ADULT 120"-LF
DYNJAA10817;
ANESTHESIA 120" ADLT-LF
DYNJAA10775;
ANESTHESIA CIRC ADULT 120-LF
DYNJAA10601;
ANESTHESIA CIRCUIT - LF
DYNJAA10261B;
ANESTHESIA CIRCUIT 120EXP-LF
DYNJAA10741A;
ANESTHESIA CIRCUIT ADULT-LF
DYNJAA0380A
DYNJAA10641
DYNJAA10706B
DYNJAA10851
DYNJAA10922B
DYNJAA10971A
DYNJAA11030A
DYNJAA11087
DYNJAA11126;
ANESTHESIA CIRCUIT-LF
DYNJAA0309A
DYNJAA0368A
DYNJAA10368A
DYNJAA10392
DYNJAA10430A
DYNJAA10547A
DYNJAA10639A
DYNJAA10656A
DYNJAA10685A
DYNJAA10685B;
CIRC 120" EXP 1 BV HMEF 3L GSL
DYNJAA10920;
EXPANDABLE CIRCUIT-LF
DYNJAA11095;
NM ANESTHESIA TURNOVER KIT
DYNJAA0433;
PROV HEALTH HEPA CIRC #2-LF
DYNJAA0360;
SPTMPX01IA CIRCUIT ADULT-LF
DYNJAA10973A;
Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Circuit, Breathing (W Connector, Adaptor, Y Piece)Distribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK
AL
AR
AZ
CA
CO
CT
FL
GA
IA
IL
IN
KS
KY
MA
ME
MI
MN
MO
MS
MT
NC
NJ
NY
OH
OR
PA
PR
RI
SC
TN
TX
WA
WI;Recall #Z-0442-2026
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 10, 2025
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits;
BREATHING KIT-LF
DYNJ10550
Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Anesthesia KitDistribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK
AL
AR
AZ
CA
CO
CT
FL
GA
IA
IL
IN
KS
KY
MA
ME
MI
MN
MO
MS
MT
NC
NJ
NY
OH
OR
PA
PR
RI
SC
TN
TX
WA
WI;Recall #Z-0443-2026
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 10, 2025
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits;
OPEN HEART SUPPLEMENT
DYNJ909116D;
VASCULAR
DYNJ909124F
Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Cardiovascular Procedure KitDistribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK
AL
AR
AZ
CA
CO
CT
FL
GA
IA
IL
IN
KS
KY
MA
ME
MI
MN
MO
MS
MT
NC
NJ
NY
OH
OR
PA
PR
RI
SC
TN
TX
WA
WI;Recall #Z-0444-2026
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 10, 2025
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits;
ENT
DYNJ909128I;
LARYNGOSCOPY - BRONCHOSCOPY
DYNJ909114D
Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Ear, Nose, And Throat Surgical TrayDistribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK
AL
AR
AZ
CA
CO
CT
FL
GA
IA
IL
IN
KS
KY
MA
ME
MI
MN
MO
MS
MT
NC
NJ
NY
OH
OR
PA
PR
RI
SC
TN
TX
WA
WI;Recall #Z-0445-2026
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 10, 2025
DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF
DYNJ909127F CYSTO
DYNJ909129D GENERAL LAPAROSCOPY-LF
Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Laparoscopy KitDistribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK
AL
AR
AZ
CA
CO
CT
FL
GA
IA
IL
IN
KS
KY
MA
ME
MI
MN
MO
MS
MT
NC
NJ
NY
OH
OR
PA
PR
RI
SC
TN
TX
WA
WI;Recall #Z-0446-2026
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 10, 2025
Convenience Kit, BREAST HERNIA
DYNJ909125G
Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Plastic Surgery And Accessories KitDistribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK
AL
AR
AZ
CA
CO
CT
FL
GA
IA
IL
IN
KS
KY
MA
ME
MI
MN
MO
MS
MT
NC
NJ
NY
OH
OR
PA
PR
RI
SC
TN
TX
WA
WI;Recall #Z-0447-2026
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 10, 2025
Convenience Kits:
DYNJ901048N ORTHO SPINE ST CLAIR
DYNJ909123F UROLOGY MINOR
DYNJ909132F SET UP
DYNJ909308D HIP PINNING
Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: General Surgery TrayDistribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK
AL
AR
AZ
CA
CO
CT
FL
GA
IA
IL
IN
KS
KY
MA
ME
MI
MN
MO
MS
MT
NC
NJ
NY
OH
OR
PA
PR
RI
SC
TN
TX
WA
WI;Recall #Z-0448-2026
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 10, 2025
Anesthesia circuit kit;
CIRCUIT,ANES,EXTENDFLEX,120,LF
DYNJAA10230
Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Filter, Bacterial, Breathing-CircuitDistribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK
AL
AR
AZ
CA
CO
CT
FL
GA
IA
IL
IN
KS
KY
MA
ME
MI
MN
MO
MS
MT
NC
NJ
NY
OH
OR
PA
PR
RI
SC
TN
TX
WA
WI;Recall #Z-0449-2026
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 10, 2025
Convenience kits;
DYNJ901046M GYN ROBOTIC
DYNJ909120D ROBOTIC PROSTATE
Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Gynecological Laparoscopic KitDistribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK
AL
AR
AZ
CA
CO
CT
FL
GA
IA
IL
IN
KS
KY
MA
ME
MI
MN
MO
MS
MT
NC
NJ
NY
OH
OR
PA
PR
RI
SC
TN
TX
WA
WI;Recall #Z-0450-2026
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 10, 2025
Convenience kits;
DYNJ901047R ORTHO SPINE FUMICH
DYNJ902308K TOTAL JOINT
DYNJ909111F DR SALMA'S SPINE
DYNJ909117G PATTERSON TOTAL JOINT
DYNJ909113F KYPHOPLASTY
Reason: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Orthopedic TrayDistribution: International distribution to Mexico and Panama. US nationwide distribution to the following: AK
AL
AR
AZ
CA
CO
CT
FL
GA
IA
IL
IN
KS
KY
MA
ME
MI
MN
MO
MS
MT
NC
NJ
NY
OH
OR
PA
PR
RI
SC
TN
TX
WA
WI;Recall #Z-0451-2026
Open, ClassifiedPosted Nov 13, 2025·Initiated Oct 6, 2025
IMMULITE 2000 Oak Mix Specific Allergen;
Siemens Material Number (SMN): 10386105;
Catalog Number: L2KUN6;
Reason: The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Root cause: Under Investigation by firm
Recalling firm: Siemens Healthcare Diagnostics, Inc.Type: System, Test, Radioallergosorbent (Rast) ImmunologicalDistribution: Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.Recall #Z-0491-2026