Zimmer · Model ATS 1500
Indicated to temporally occlude blood flow in a patient’s extremities during surgical procedures on those extremities Tourniquets have been found useful in producing a bloodless operating field in surgical procedures involving the extremities. Line Voltage Rage: 105-130 VAC, 50/60 Hz Line Current: 485 mA RMS Max., 300 mA RMS nominal Input Power: 58 W max, 36 W nominal Power Cord Type: SJT, AWG 18, 13 ft. 2 in (4.01m) Power Plug Hospital grade, 3 prong straight blade, 15A Line Protection: 0.5A, 250V, 5 x 20 mm SLO-BLO fuse Cuff Pressure Range: 50-475 mmHg Pressure Accuracy: ± 5 mmHg (50-475 mmHg) Pressure Regulation: ± 6mm of set point Maximum Pressure: 475 mmHg (Nominal operation), 600 mmHg (internally limited) Time Alarm Set Ranges: 0-240 minutes; 1 minute increments Timer Accuracy: 0.25% ± 1 second.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pressure droops during a case, will not inflate to setpoint, leak alarm.
Root cause: Internal pump / compressor wear, a solenoid-valve fault, or a leak in the tubing / cuff / connectors.
Symptoms: Displayed pressure differs from an independent gauge - patient-safety critical.
Root cause: Pressure-transducer drift or a calibration fault.
Symptoms: One channel will not regulate independently, cross-talk between channels.
Root cause: Channel solenoid / transducer or a manifold fault.
Symptoms: Inflation-time alarm not sounding, timer wrong.
Root cause: Alarm / timer board fault - a safety function.
Symptoms: Leak at the connector, cuff will not hold.
Root cause: Worn tubing, connectors, or a damaged cuff.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Verify against a calibrated gauge after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Internal pump / compressor | Major service part | $$$ | Difficult |
| Solenoid / regulator valve | Major service part | $$ | Moderate |
| Pressure transducer Verify against a calibrated gauge after replacement. | Periodic replacement | $$ | Moderate |
| Connecting tubing / connectors | Frequent wear item | $ | Easy |
| Alarm / control board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.