Vitrolife Sweden AB · Model 14220
Ultrasound Transducer Cover.
Intended use: A sheath intended to be used as a physical barrier for protection against the effects of environmental exposure (e.g., body fluids, gliding creams) and/or to maintain the required hygienic level of an ultrasound imaging system transducer (probe) used in natural body orifices (e.g., vagina, rectum, oesophagus) during an ultrasound examination; it has no additional functionality. This is a single-use device.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Black vertical band in the image, halved sensitivity, artefact that stays with the probe on any port.
Root cause: Cracked or delaminated piezo elements, or broken cable conductors into the array.
Symptoms: Near-field haze, bubbles under the lens, visible crack or a bite mark (TEE).
Root cause: Impact, aggressive disinfectant, or age; the acoustic lens has lifted from the array.
Symptoms: Image comes and goes when the cable or connector is flexed.
Root cause: Broken conductors at the connector or at the probe strain relief.
Symptoms: Fails the leakage / insulation test; unsafe for patient contact (critical for intracavitary and TEE probes).
Root cause: Fluid ingress or a lens breach compromising patient isolation.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Element or cable damage is usually an OEM/third-party repair, not a field fix.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transducer (exchange / repair) Element or cable damage is usually an OEM/third-party repair, not a field fix. | Periodic replacement | $$$$ | Fairly easy |
| Connector pins / housing | Major service part | $$ | Moderate |
| Acoustic lens / membrane | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio