Thermo Scientific · Model Jouan KR4i
Powerful brushless motor: Designed for continuous processing. Blood Banking: 12 x Quin/Quad/Triple bags (buckets 2,000 ml), 8 x Quin/Quad bags with in-line Filter (patented liner, buckets 2,160 ml) or 6 x Quad/Triple bags. Processing: 4 x 1,600 ml, 6 x 1000 ml. Pathology and Research: Accepts up to 504 samples and all types of sample container. Analyzer Sample Preparation: Accepts huge quantities of racks from analyzers of many brands for RIA, Hematology, Biochemistry. Large choice of accessories: Offers three swing-out rotors, a wide selection of accessories and an unparalleled microplate capacity providing the ablity to process up to 48 standard microplates or 12 to 18 deepwell microplates in a single run. Intuitive controls: VIDEOsetTM graphic control system with pictures and messages. Careful sample handling: Liners enable blood bags to be inserted and removed without force. High precision: Microprocessor control of all parameters. Lifetime reproducibility: Guaranteed run to run by automatic High Reproducibility Factor (integral). Ideal for platelet production. True RCF: Using programmed radius allowing for use of adaptors. Ideal for process development: 24 program memory. Accurate temperature control: Patented temperature sensor and software independent of the bowl for better sample temperature control. Convenient: Can run without rotor lid for instant access to samples. Maintenance-free motor: Quiet, brushless induction motor, designed and manufactured by Thermo. Absolute safety: Sealed buckets and microplate carriers certified to IEC 1010-2-020 for biocontainment by C.A.M.R.* Porton Down. Imbalance detection & shutdown, guard barrier, dual lid interlock, current and temperature limitation.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Run ends early / late, or RPM does not match the setting.
Root cause: Tacho / speed-sensor drift or a control-board fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Motor brushes (older units) | Periodic replacement | $ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.