Tanita · Model BF-350
Rugged design with a proven track record of reliability. Suitable for automated applications within a kiosk setting. Within 5% of DEXA, at a fraction of the cost. EMR ready with straightforward digital outputs. 440 pound weight capacity suitable for all but the very obese patients. Wireless Options include ANT and WiFi.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Reading disagrees with certified test weights, drifts, or fails calibration.
Root cause: Load-cell drift or damage, a binding platform / linkage, or a calibration fault.
Symptoms: Reading will not settle or return to zero, jumps around.
Root cause: Something fouling the platform, a damaged load cell, or electrical noise / a loose connector.
Symptoms: Missing segments, unresponsive keys, error code on power-up.
Root cause: Display or keypad membrane fault, or a main-board fault.
Symptoms: Will not power on, short battery life, only works on mains.
Root cause: Aged battery or a power / adapter fault.
Symptoms: Wobble, worn castors, loose seat or handrail, folding mechanism stiff.
Root cause: Worn castors, hardware, or the folding mechanism.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate with certified weights after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Load cell(s) Recalibrate with certified weights after replacement. | Major service part | $$ | Moderate |
| Display / main board | Major service part | $$ | Moderate |
| Keypad membrane | Periodic replacement | $ | Fairly easy |
| Battery / power adapter | Periodic replacement | $ | Easy |
| Castors / seat / handrail hardware | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.