Stryker · Model Visum LED
Stryker's Visum LED surgical light.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: One or more heads / LED segments out, dim head, uneven or discoloured light field.
Root cause: Halogen lamp at end of life or a failed LED module / driver, a lamp-changeover fault, or a power-supply fault.
Symptoms: Light head drifts up or down, will not stay positioned.
Root cause: Spring-arm counterbalance out of adjustment or worn friction brakes.
Symptoms: Brightness or spot size will not adjust from the wall panel or sterile handle.
Root cause: Control PCB, the wall-plate keypad, or the handle switch.
Symptoms: Handle loose or not detected.
Root cause: Worn handle mount / contacts.
Symptoms: Whole head dark, flicker, or will not power on.
Root cause: Failed lamp / LED power supply.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Light-head lamp / LED module | Frequent wear item | $$ | Fairly easy |
| Lamp / LED power supply | Major service part | $$ | Moderate |
| Spring-arm counterbalance / friction-brake kit | Major service part | $$$ | Difficult |
| Control PCB / wall-plate keypad | Major service part | $$ | Moderate |
| Sterilizable handle | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
There is a potential the joint in the suspension of the device is insufficiently assembled.
The potential hazard that can arise for the combination of the missing screws and the overloading by the end user is insufficient mount force that causes the equipment to fall.
Potential incomplete final weld on the drop tube flange.
Stryker has become aware that there is a low likelihood of incomplete engagement of the Spring Arm Circlip during service or replacement of the Spring Arm. If the Spring Arm Circlip is not fully seated, the Spring Arm could separate and fall along with the surgical lights or monitors mounted to the end of the Spring Arm. To date, there have been no serious injuries reported from the incomplete e
Stryker has identified a potential structural issue with the Visum LED surgical light(s). During a complaint investigation, Stryker became aware that there was a possibility that incorrect assembly could result in the light head beginning to separate from the suspension arm. In the event the light separates and falls from the arm, it could lead to serious injury.
Installation records for the SwitchPoint Infinity, Booms, Lights, and Flat Panel systems are incomplete or inadequate.
Installation records are incomplete.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.