Stryker · Model Prime TC
Big wheel maneuverability Moveable, but not removable components Yellow activation points make operation intuitive for all users One-touch central brake
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Won't hold position, rolls when locked.
Root cause: Brake pad wear or linkage failure.
Symptoms: Hard to push straight, wobbles or drags.
Root cause: Bearing wear or a flat-spotted wheel.
Symptoms: Visible crack or looseness at a joint.
Root cause: Repeated load cycling.
Symptoms: Torn or cracked cover, exposed padding.
Root cause: Cleaning-chemical and UV degradation, routine wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace on visible wear or tear.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Brake / wheel-lock mechanism | Periodic replacement | $$ | Fairly easy |
| Upholstery / seat or surface cover Consumable - replace on visible wear or tear. | Periodic replacement | $ | Easy |
| Frame hardware (bolts, pins, bushings) | Major service part | $ | Fairly easy |
| Caster wheel assembly | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance additives as the specified oils and therefore its use increases the potential for wear which may cause a leak event or jack drift. Testing was done to evaluate the jacks with incorrect oil and leak events we
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.