Multi-Lesioning: Handles up to 4 lesions simultaneously and with independent control Offers a multiple start option as well as an individual start option to deliver RF energy Performs monopolar and bipolar procedures at the same time Reduces overall procedure time Parallel Bipolar: Allows 2 parallel bipolar procedures to be performed simultaneously Features an option to perform thermal and pulsed procedures at the same time Intradiscal Lesioning (IDL): Offers pre-programmed time and temperature step profiles Allows time and temperature adjustments on the fly.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Lesion not forming, "high impedance" or "open circuit", output shuts down.
Root cause: Poor electrode contact or a dry field, a damaged RF cable or probe, or a grounding-pad fault.
Symptoms: Tip temperature reads implausibly, will not reach or hold the set lesion temperature.
Root cause: Damaged thermocouple in the cannula / probe, or a temperature-control fault.
Symptoms: Pad-contact-quality alarm, output inhibited.
Root cause: Poorly applied or dried dispersive pad, or a pad-cable fault.
Symptoms: Stimulation test gives no or erratic response.
Root cause: Stimulation output stage or cable / probe fault.
Symptoms: Unresponsive display, will not power on.
Root cause: Display, power supply, or control board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reusable RF connector cable | Periodic replacement | $$ | Easy |
| RF probes / cannulae with thermocouple | Frequent wear item | $$ | Easy |
| Dispersive (grounding) pads | Consumable | $ | Easy |
| Footswitch | Periodic replacement | $$ | Fairly easy |
| Power supply / control board | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.