STERIS · Model HarmonyAIR M-Series
Fat beam illumination - more uniform light disrtibution across the spot. Pure white light - 4400K +/- 300K color temperature. Energy efficient - consumes 1/3 the heat of previous LED technology (M-5 model only). 84 overlapping beams of light virtually eliminate shadows from the surgeons head and hands (M-5 model only). Better tissue discrimination with a CRI of up to 96. Better discrimination of deep, saturated reds with an R9 value of up to 98. An open architecture design accommodates any HD signal, regardless of the manufacturer or signal type. Grows with you - add lightheads, ambient light, monitors, radiation shield and in light camera. Add Harmony ConnectPoint Video Connectivity System for video integration without costly or involved renovations. Convert general rooms to specialty suites quickly and easily. Flat panel, easy position monitor yokes holds up to 32” monitors. Suspension systems are American-made and feature STERIS-exclusive design.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: One or more heads / LED segments out, dim head, uneven or discoloured light field.
Root cause: Halogen lamp at end of life or a failed LED module / driver, a lamp-changeover fault, or a power-supply fault.
Symptoms: Light head drifts up or down, will not stay positioned.
Root cause: Spring-arm counterbalance out of adjustment or worn friction brakes.
Symptoms: Brightness or spot size will not adjust from the wall panel or sterile handle.
Root cause: Control PCB, the wall-plate keypad, or the handle switch.
Symptoms: Handle loose or not detected.
Root cause: Worn handle mount / contacts.
Symptoms: Whole head dark, flicker, or will not power on.
Root cause: Failed lamp / LED power supply.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Light-head lamp / LED module | Frequent wear item | $$ | Fairly easy |
| Lamp / LED power supply | Major service part | $$ | Moderate |
| Spring-arm counterbalance / friction-brake kit | Major service part | $$$ | Difficult |
| Control PCB / wall-plate keypad | Major service part | $$ | Moderate |
| Sterilizable handle | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio