Spacelabs · Model Ultraview SL Command Module
Ultraview SL Command Module 91496 The Ultraview SL Command Module is the core of the SL patient monitoring system, providing the processing power for all basic physiologic parameters. You can select from a variety of configurations to suit the monitoring needs of specific patients or care units in the hospital. In addition, three levels of arrhythmia monitoring as well as diagnostic ECG analysis and reports with or without measurement and interpretation are available. You can also add ST-segment analysis and event review or Varitrend 4 for event review of neonatal respiration, heart rate and SpO2. You have a choice of technologies: Spacelabs, Masimo SET, or Nellcor OxiMax.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.
The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW Error, or No Data.
Modules resetting. This failure mode happens when the Spacelabs Command Module is configured with the Masimo SpO2 option (-M), the Spacelabs Respiration option (-R) and the Masimo SpO2 PCBA, PN: 010-1136-02. With these two options and the Masimo SpO2 PCBA present, the module may experience random resets.
Ultraview SL Command Modules which were upgraded with the Masimo SpO2 Option, Model 91496-M was affected. Customer reported that during the time that the module is resetting (~10 seconds), monitoring of all parameters will be suspended. Upon completion of this sequence, alarms will be reset to their default values.
Spacelabs Healthcare XPREZZON Bedside Monitors, Model 91393, are recalled because the firm has received multiple reports of XPREZZON monitors failing to boot up or returning to factory default configuration settings following power on or reset.
With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.
A software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms.
There is a potential failure of the Integrated Module Housing causing the patient module parameters to drop off the monitor display or the patient module will fail to sign on when inserted into the housing.
The screw used to connect the external power supply to the Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, may damage an internal capacitor. The damage may result in an inability of the monitor to power on, power off, or involuntarily shut down.
NIBP pump inside the Ultraview SL Command Module, Model 91496, became loose. The pump is suspended by the air hoses and electrical connections, but there can be enough movement to damage internal components and cause the module to fail.
One Spacelabs Medical Ultraview SL Command Module, Model 91496, was removed from customer due to software error.
The circuit that supplies backup power to the modules in the Medical Patient Monitor, Model 91387, is not working. If there is a power interruption, the three-minute backup may not be operational and customers may lose any changes they have made to the alarm settings.
Monitor may reset repeatedly which can result in a loss of patient monitoring for several minutes. Cause is a software problem.
Potential for module to fail to obtain a NIBP reading on hypertension patients.
Potential for modules to reset, alarms returning to default settings, invasive pressure losing its labels and zero calibration and Sp02 alarm limits being frozen. This can result in delay in treatment.
The transducer used to compensate for pressure changes was mounted backwards on the printed circuit board assembly. The module will not compensate for pressure changes and CO2 readings will be inaccurate at higher altitudes.