Spacelabs · Model 90467 Adult NIBP/SpO2 Module
Spacelabs 90467 Adult NIBP/SpO2 Module
No documents available for this device yet.
Synced live from the FDA openFDA API.
The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.
Spacelabs has received multiple complaints reporting telemetry beds dropping off the Spacelabs Central Station resulting in a loss of monitoring.
The firm received reports of telemetry SpO2 numerics dropping off the Xhibit Central display. Desaturation, high, and low limit alarms work normally.
The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW Error, or No Data.
Modules resetting. This failure mode happens when the Spacelabs Command Module is configured with the Masimo SpO2 option (-M), the Spacelabs Respiration option (-R) and the Masimo SpO2 PCBA, PN: 010-1136-02. With these two options and the Masimo SpO2 PCBA present, the module may experience random resets.
Ultraview SL Command Modules which were upgraded with the Masimo SpO2 Option, Model 91496-M was affected. Customer reported that during the time that the module is resetting (~10 seconds), monitoring of all parameters will be suspended. Upon completion of this sequence, alarms will be reset to their default values.
The transmitter display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions. This is an expansion of recall Z-0871-2014 because additional serial numbers were identified as being part of this recall action as part of the firm's retrospective analysis of all recalls since 2011.
The AriaTele Model 96281 with SpO2 monitoring Option C is recalled because the AriaTele (transmitter) display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions.
The screw used to connect the external power supply to the Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, may damage an internal capacitor. The damage may result in an inability of the monitor to power on, power off, or involuntarily shut down.
NIBP pump inside the Ultraview SL Command Module, Model 91496, became loose. The pump is suspended by the air hoses and electrical connections, but there can be enough movement to damage internal components and cause the module to fail.
One Spacelabs Medical Ultraview SL Command Module, Model 91496, was removed from customer due to software error.
The monitor has a 3 second delay in the waveforms and respiration alarm; the delay is not disclosed in the IFU.
There have been reports that the Spacelabs Medical Vital Signs Monitor, Model 91220, will not power up with AC power and will no longer charge the batteries.
Potential for module to fail to obtain a NIBP reading on hypertension patients.