Sorvall · Model RC4
Floor centrifuge Process the optimal capacity for small to medium size blood processing centers with the ability to spin 4 or 8 blood bags in a wide range of bag configurations when using the Thermo Scientific™ Sorvall™ RC4 Low-Speed Floor Model Centrifuge, a flexible refrigerated centrifuge for blood banking and clinical applications. Using flexible accessory systems, it spins everything from doubles to 500mL quints with in-line leukoreduction filters. A versatile performer, the RC-4 is in a class of its own.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid not locked, or a drive / control fault.
Symptoms: Slow spin-up, rumble or vibration.
Root cause: Load imbalance, worn motor bearings, worn drive coupling, or (brushed motor) worn brushes.
Symptoms: Lid-lock error, lid will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still in progress.
Symptoms: Chamber will not reach set temperature.
Root cause: Compressor, condenser, or a sealed-system fault.
Symptoms: RCF / RPM or chamber temperature does not match the setting.
Root cause: Tacho / speed-sensor or temperature-sensor drift.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Motor bearings / brushes | Major service part | $$ | Difficult |
| Drive coupling / belt | Major service part | $$ | Moderate |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed / temperature sensor | Major service part | $$ | Moderate |
| Refrigeration components (refrigerated models) | Major service part | $$$$ | Expert-level |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio