Sorvall · Model Primo Series
Primo centrifuges, with their applications-oriented accessories, are proven favorites in clinical, molecular biology, cell biology and small medical laboratories. Featuring an automatic rotor recognition system and electronic imbalance protection, whisper quiet Primo instruments accept fixed-angle, swinging bucket and microliter rotors. The innovative Highconic® 6 x 50 ml fixed-angle rotor allows the centrifugation of conical cell culture vessels at up to 10,000 x g, while a high-performance 24-place microliter rotor rapidly accelerates 1.5 or 2 ml tubes to a maximum RCF of 21,885 x g. Both rotors are autoclavable to 121°C - and along with the accessory 4 x 100 ml swinging bucket rotor, are certified as providing biocontainment for biologically safe centrifugation. Process control is via the EASYcontrol™ II panel, clearly displaying all operational and run parameters. Time, speed and RCF values are set using a touch pad, with 2 acceleration and 9 braking curves available for sensitive samples. A quick-run key is also available for short spins.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Run ends early / late, or RPM does not match the setting.
Root cause: Tacho / speed-sensor drift or a control-board fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Motor brushes (older units) | Periodic replacement | $ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.