OPHTHALMOLOGY SLIT LAMP MICROSCOPE This is a slit lamp microscope used by ophthalmologists in the field of ophthalmology for eye examination. The magnification is selectable between 6x, 10x, 16x, 25x, and 40x. The included eyepiece pair is 12.5x. It comes with optional R and T type applanation tonometers as well as an optional measuring eyepiece. Another option is a beam splitter, ccd camera adapter, and ccd camera for taking the video image to a cctv monitor for all in the room to see, or for the ophthalmic examination to be videotaped. Another useful option is an extra monocular teaching tube, allowing a third observer/assistant to view the ophthalmic eye examination simultaneously. The lens is made from high quality imported German glass for reduction in lens aberration. All optics are multi-coated and include mildew-proof and antireflection coatings. This slitlamp microscope is made in a US FDA (Food and Drug Administration) registered manufacturing facility as well as has CE certification.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No or dim slit, uneven illumination, colour shift.
Root cause: Halogen / LED source at end of life, a bulb-holder / contact fault, or a rheostat fault.
Symptoms: Slit width, length, angle or filter selection will not change or sticks.
Root cause: Worn slit mechanism or filter turret.
Symptoms: Coarse or fine movement binds, drifts, or the locking screw is ineffective.
Root cause: Dry or worn base slide / bearings, or a loose locking screw.
Symptoms: Haze, spots, or reduced image quality through the oculars.
Root cause: Dust or fingerprints on the objective, oculars, or mirrors.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Illumination source (halogen / LED) | Frequent wear item | $ | Fairly easy |
| Rheostat / illumination control | Major service part | $$ | Moderate |
| Slit mechanism / filter turret | Major service part | $$$ | Difficult |
| Base-slide bearings / glide plate | Major service part | $$ | Moderate |
| Mains fuse | Major service part | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.