Skytron · Model Stellar
Brighter - Soft white, high intensity lighting @ 130,000 lux (ST23) or 160,000 lux (ST29) per lighthead. Direct Focus Control - Surgeons can directly control focus and depth of field adjustment without having to move the lighthead, while retaining maximum intensity at any distance. Non-Sterile Focus Control - A non-sterile focus control knob and four non-sterile positioning handles for the circulating nurse are standard. Precise, Shadow Free Optics - Advanced VSRD (Vertical Segmented Reflector Design) Optics provide precise, shadow free illumination and a uniform, adjustable 30” depth of field at any distance. Cool - Revolutionary coated reflectors reflect infrared heat to the rear of the light-head, providing the optimum balance between bright light and cool illumination.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: One or more heads / LED segments out, dim head, uneven or discoloured light field.
Root cause: Halogen lamp at end of life or a failed LED module / driver, a lamp-changeover fault, or a power-supply fault.
Symptoms: Light head drifts up or down, will not stay positioned.
Root cause: Spring-arm counterbalance out of adjustment or worn friction brakes.
Symptoms: Brightness or spot size will not adjust from the wall panel or sterile handle.
Root cause: Control PCB, the wall-plate keypad, or the handle switch.
Symptoms: Handle loose or not detected.
Root cause: Worn handle mount / contacts.
Symptoms: Whole head dark, flicker, or will not power on.
Root cause: Failed lamp / LED power supply.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Light-head lamp / LED module | Frequent wear item | $$ | Fairly easy |
| Lamp / LED power supply | Major service part | $$ | Moderate |
| Spring-arm counterbalance / friction-brake kit | Major service part | $$$ | Difficult |
| Control PCB / wall-plate keypad | Major service part | $$ | Moderate |
| Sterilizable handle | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.