Siemens Healthineers · Model Acuson 17L5 Linear Array
Frequencies: 15.0, 12.0, 8.0 MHz. NTHI Frequency: H14.0 MHz. Color Doppler Frequencies: 14.0, 12.0, 10.0, 8.5, 7.0 MHz. Spectral Doppler Frequency: 7.0 MHz. High Resolution Color Flow. (HRCF) Frequencies: 14.0, 10.0, 7.0 MHz. Cadence™ Contrast Pulse. Sequencing (CPS) Frequencies: P14.0, P10.0, P8.0, P7.0 MHz. Imaging Formats: Linear and Vector™ wide-view imaging format. Array Footprint: 51.2 mm. Maximum Display Depth: 80 mm.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Persistent dark bands or a missing sector that do not move with the anatomy.
Root cause: Failed piezoelectric elements in the array.
Symptoms: Image drops, flickers or is noisy when the cable or strain relief is moved.
Root cause: Broken conductors in the transducer cable or at the connector.
Symptoms: Haze, comet artefacts, reduced penetration, a visible cut or bubble under the lens.
Root cause: Lens delamination or a puncture; drop damage.
Symptoms: System does not detect the probe or shows the wrong probe; bent or corroded pins.
Root cause: Damaged connector pins or a wet / dirty connector.
Symptoms: (TEE / endocavity) fails a pre-use leak test, tip articulation loose or locked, bite-mark damage.
Root cause: Worn articulation cables / seals or a housing breach.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Manufacturer / authorised repair only.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transducer cable Manufacturer / authorised repair only. | Periodic replacement | $$$ | Difficult |
| System connector assembly | Major service part | $$$ | Difficult |
| Acoustic lens / membrane | Periodic replacement | $$$ | Expert-level |
| Strain-relief boot | Periodic replacement | $ | Moderate |
| Articulation / seal service kit (TEE / endocavity) | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio