Saturn · Model Evo Color
Spontaneous or manual ventilation • Manual/Spontaneous Ventilation Mode + VoluProtect Volume-controlled ventilation • CMV - Controlled Mandatory Ventilation + BaroProtect • SIMV - Synchronised Intermittent Mandatory Ventilation with built-in VOLUME SUPPORT (VS) + BaroProtect Pressure-controlled ventilation • PCV - Pressure-Controlled Ventilation with built-in PRESSURE SUPPORT (PS) With adequately set pressure trigger PCV becomes: PC-SIMV - Synchronised Intermittent Pressure-Controlled Ventilation with built-in PRESSURE SUPPORT (PS) • PSV - Pressure Support Ventilation with Apnea Backup The Saturn Evo Color anaesthesia workstation also comprises • 10” full-colour touchscreen with control knob • ORC Rotameter (with 25% oxygen safeguard and oxygen alarm) • Tripple Selectatec® vaporiser mounting system • APL arm with Y-piece connector and bag hook • Extendable writing shelf • Patient tubing holder • Three drawers • Advanced absorber block (CO2 bypass) with soda lime canister • Common gas outlet • Integrated suction system with pressure gauge • Three gas pressure gauges (O2, N2O, air) • Three cylinder holders at back • Anti-static castors with brake • Integrated Gasovac for active scavenging system
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Delivered agent concentration off, low/high agent alarm.
Root cause: Worn bellows or O-ring; agent contamination.
Symptoms: O2 low-pressure alarm, inability to ventilate.
Root cause: Pipeline disconnect or failed O2 flush valve.
Symptoms: Low minute volume, volume loss, leak alarm.
Root cause: Loose connections, torn bellows, cracked hose.
Symptoms: Rising EtCO2, patient rebreathing.
Root cause: CO2 absorbent saturated.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable, replace on schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Breathing circuit / bellows | Periodic replacement | $$ | Fairly easy |
| CO2 absorber canister Consumable, replace on schedule. | Consumable | $ | Easy |
| Scavenger interface & hose | Periodic replacement | $$ | Fairly easy |
| O2 pressure regulator | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →