Sartorius · Model SIGMA 3-18 and 3-18K
SIGMA 3-18. Universal table top centrifuge; speed range up to 16.000 rpm. Operation with various fix-angle and swing-out rotors available. Maximum capacity of 4 x 250 ml down to microtubes with different adapters in the same bucket. Maintenance-free induction drive motor. A microcontroller controls speed, time, temperature and gravitational field. 10 quadratic and 10 linear preset deceleration and acceleration curves. Any curve can freely be programmed and displayed grafically. Suitable for low speeds e.g. for cell sedimentation, accuracy +/– 1 rpm. Magnetic rotor identification prevents the rotor from overspeeding. Stainless steel bowl. Imbalance switch. Produced according to national and international safety regulations. Motorized lid lock SIGMA 3-18 K. Technically largely identical to 3-18 but with built-in refrigeration system. Suitable for low speeds e.g. for all sedimentation. Higher speed for faster separations. Possibility of precooling the rotors during standstill. Refrigerated CFC-free (R 134 a). Suitable for sedimentation of temperature sensitive substances.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Run ends early / late, or RPM does not match the setting.
Root cause: Tacho / speed-sensor drift or a control-board fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Motor brushes (older units) | Periodic replacement | $ | Moderate |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio