Samsung · Model WS80A Elite
High Resolution, Premium Women’s Health Ultrasound Hybrid Digital Beamforming Technology and S-Vue Probe Technology Deliver Exceptional Image Clarity 5D Advanced Applications Available Including 5D Heart, CNS, LB, NT and Follicle to Streamline Workflow Fetal Realistic Vue(FRV) Offers 3D/4D Rendering with Excellent Detail and Realistic Depth Penetration 5D Heart (Fetal Heart Examination) 5D Heart is an innovative application which generates and displays 9 standard fetal echocardiography views by applying intelligent navigation technology to volume datasets of the fetal heart. 5D heart has an intuitive workflow that can simplify examination of the fetal heart and reduce operator dependency. [u]5D CNS (Fetal Brain Measurement)[/u] Using intelligent navigation, 5D CNS provides six measurements (BPD, HC, OFD, Cerebellum, Cisterna Magna, Atria of lateral ventricle) from three transverse views of the fetal brain to help facilitate measurement reproducibility and streamline workflow. [u]5D LB (Fetal Long Bone Detection)[/u] 5D LB efficiently locates, displays and measures fetal long bones (Radius, Ulna, Humerus, Femur, Tibia and Fibula) from within the 3D dataset, streamlining workflow while helping to facilitate measurement reproducibility. [u]5D NT (Nuchal Translucency Measurement)[/u] 5D NT intelligent navigation helps to simplify the NT measurement process by locating the precise mid-sagittal plane in a single step. Semi-automatic measurements are then performed for improved NT accuracy and reduce examination time. [u]Designed For Convenience[/u] Streamline examination with 23" Full HD LED monitor, 10.1" touch screen and height-adjustable control panel. The adjustable gel warmer maintains a comfortable gel temperature from 80°F to 102°F. Two endocavity transducer holders are conveniently placed, one side-mounted and one behind the console.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio