Samsung · Model Accuvix A30
Ergonomic Design: flexible controls, articulated monitor arms, all-new user interface, central locking (conveniently locking with foot control) Dynamic MR Plus: The Dynamic MR Plus is a 2D image feature which enhances detection and contrast resolution while decreasing speckle echoes. Enhanced DPDI: The Accuvix A30 with its enhanced Directional Power Doppler Imaging (DPDI) detects peripheral blood vessels. Thyroid ElastoScan: The extreme sensitivity of Thyroid ElastoScan™ enables early detection of tumors by analysing changes or abnormalities in thyroid tissue. Breast and Cervix ElastoScan: ElastoScan is a highly accurate, cutting-edge technology that assists with the early detection and differentiation of malignant tumors and various other diseases. Volume NT & IT (NT: Nuchal Translucency, IT: Intracranial Translucency) is an innovative 3D technology that captures more detailed and comprehensive views of the mid-sagittal plane. HD Volume Imaging: HD Volume Imaging (HDVI) uses a special filtering technique to remove unwanted noise, while simultaneously improving the visualisation of the edges and small structures in volume data. Advanced 3D Feature: VSI, SFVI, Smooth Cut, FAD. More comprehensive and realistic images which lead to increased accuracy in scans and diagnosis. Colour Opt Flow (enables colour and power imaging of blood flow and is compatible with Doppler sensitivity settings). Hybrid Beamforming Engine (enables more indepth and detailed scanning). HD ADVR (enables simultaneous scanning and recording, and allows users to choose desired recording areas). 54.61cm (21.5) Full HD LED Monitor for sharper, easier-to-view images. EZ Exam (transforms frequently used step-by-step exams into one single, streamlined procedure to offer enhanced usability).
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio