REXMED · Model ROL-501C
MOD. ROL-501C, AC 100~240V, Surgical Lamp, HS CODE: 940520, UNCCS CODE: 486304, UMDNS CODE: 12282, GMDN CODE: 32244, Main construction: cast aluminium and powder coated steel, Ceiling mounted type, Single dome surgical lamp, Lamp head: 62cm with 4,760 lenses / transparent lenses polycarbonate reflector cover, Reflector: 50cm aluminium 160 layers reflector and 8pcs filter, Bulb specification: OSRAM 24V 150W x 3pcs (1 primary and 2 backup X 1set), Bulb life time: 3,000 hrs ±10%, Colour Temperature: 3,800/4,300/4,500K ±200K, CRI: 94 ±3, Working distance: 60~180cm, Depth of light (L1+L2): 120cm, Intensity: 0~80,000/100,000/150,000 Lux/1m, Focus: 15~25cm by sterilizable handle, light field: d10=15cm, d50=8cm, Remaining intensity / EC with one mask: 23%, with two masks: 48%, with one tube: 97.5%, with one mask and one tube: 23%, with two masks and one tube: 48%, Swivel radius: 85+85=170cm, Lamp head elevation: 105cm 15°up/75°down, Lamp head rotation: 360°, Controller: ON/OFF switch, intensity 0~100% by 1% and 10% dimmer control on copula and infrared remote controller, Auto switch over to spare bulb < 0.1sec when main bulb failure: Yes, Surgeon head temperature increase < 2°C, Radiation energy: < 3mW/m².lux, W/S: Light bulb X 1pc, I.R. remoter controller X 1pc, Sterilizable handle X 2pcs.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: One or more heads / LED segments out, dim head, uneven or discoloured light field.
Root cause: Halogen lamp at end of life or a failed LED module / driver, a lamp-changeover fault, or a power-supply fault.
Symptoms: Light head drifts up or down, will not stay positioned.
Root cause: Spring-arm counterbalance out of adjustment or worn friction brakes.
Symptoms: Brightness or spot size will not adjust from the wall panel or sterile handle.
Root cause: Control PCB, the wall-plate keypad, or the handle switch.
Symptoms: Handle loose or not detected.
Root cause: Worn handle mount / contacts.
Symptoms: Whole head dark, flicker, or will not power on.
Root cause: Failed lamp / LED power supply.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Light-head lamp / LED module | Frequent wear item | $$ | Fairly easy |
| Lamp / LED power supply | Major service part | $$ | Moderate |
| Spring-arm counterbalance / friction-brake kit | Major service part | $$$ | Difficult |
| Control PCB / wall-plate keypad | Major service part | $$ | Moderate |
| Sterilizable handle | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio