Radiometer · Model ABL700
• Unique technology eliminates the interfering effects some substances have on the accuracy of blood gas, oximetry, glucose, and lactate measurements • Save time with automation of work processes • Auto Check quality control module automates and improves quality control procedures • Remote and automated management of patient and analyzer data from a central location through RADIANCE STAT analyzer management system software • Automatic accession number request to LIS (Laboratory Information System) • Flexible setup options to fit different needs • Simplify your routines • Simple and intuitive sample analysis procedures • Online help function with interactive video tutorials
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Aspiration error, short sample, carryover.
Root cause: Clot/crystal obstruction.
Symptoms: Vibration, loud noise, stop error.
Root cause: Worn bearing or unbalanced rotor.
Symptoms: Incubator out of range.
Root cause: Failed heater/fan or sensor.
Symptoms: Low signal, wavelength drift.
Root cause: Lamp/LED or monochromator aging.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Frequent R&R on high-throughput analyzers.
Consumable, replace proactively.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample aspiration probe Frequent R&R on high-throughput analyzers. | Periodic replacement | $$$ | Moderate |
| Peristaltic pump tubing Consumable, replace proactively. | Frequent wear item | $ | Easy |
| Xenon / LED light source | Periodic replacement | $$$ | Moderate |
| Centrifuge rotor & bearings | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t
Based on internal testing and user/customer complaint reports received, the firm determined that wrinkles may arise on the outer membrane of the calcium membrane units upon installation. Retention of fluids between the wrinkles can cause a positive bias resulting in calcium measuring errors. This situation poses a risk to the safety of patients, particularly when the measurement results are used
Reported calibration errors are not displayed on the analyzers parameter bar screen or on the patient results when the calibration number is between 32768 - 65535 and 98303 - 131070. This may cause the operator to perform measurements and obtain questionable results.
An increased failure rate was observed with the yellow rubber tubing. The tubing was found to be leaking blood and solutions after distribution.
Calibration solution for ABL700 Blood Gas Analyzers is labeled with an incorrect bar code. The bar code identified on the Cal Solution 2 product is actually the bar code for the rinse solution.
The CI membrane units generate too low values for status and sensitivity during calibration.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →