Pedigo · Model ivNow
Modular Thermal Fluid Warming ivNow modular fluid warmers put warmed fluid at your fingertips where you need it, when you need it. ivNow fluid warmers are small, convenient units that can be placed in a number of different combinations and environments. Countertop, wall mounted and mobile units are all available depending on your particular requirements. ivNow quickly warms and maintains safe temperatures of intravenous and irrigation fluids. The specially contoured warming surface cradles one (1) liter solution bags. Each cavity is controlled by an individual control and L.E.D. display. A sensor in the heating plate detects the presence of a bag and engages the heating mechanism to quickly warm the fluid. Two temperature sensors continuously read the temperature of the fluid bag to engage the heater as necessary to maintain the fluid temperature within +0/-2°C (+0/-3ºF) of the set point temperature. ivNow warmers ensure your fluids are kept at the proper temperature and will always be immediately available at or close to body temperature rather than room temperature without the added expense and set-up time required with in-line warming systems.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Blanket / fluid cabinet below (or above) setpoint, temperature alarm.
Root cause: Heater element, a temperature sensor / controller fault, a door / gasket leak, or a circulation-fan fault.
Symptoms: Cabinet stops heating, over-temp alarm, will not restart.
Root cause: Controller / sensor fault driving the heater high, or a failed independent over-temp thermostat.
Symptoms: Door will not seal, heat loss, temperature swings when opened.
Root cause: Worn door gasket or a sagging hinge.
Symptoms: Uneven temperature top to bottom, fan noise or stopped.
Root cause: Worn circulation fan.
Symptoms: Blank or frozen display, setpoint will not change.
Root cause: Controller or display fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater element | Major service part | $$ | Difficult |
| Temperature sensor / over-temp thermostat | Major service part | $$ | Moderate |
| Door gasket | Periodic replacement | $$ | Fairly easy |
| Circulation fan | Major service part | $$ | Moderate |
| Controller / display | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.