Olympus · Model CHF-CB30L\S
Bright, high resolution images - The CHF-CB30L and CHF-CB30S choledochofiberscopes are 1.8 times brighter than the previous generation CHF-CB20 and transmit a bright, clear, high resolution image that ensures a high diagnostic performance and an easy removal of stones in the biliary system. Ultra Slim Diameter for easy insertion into the cystic and common bile duct - The CHF-CB30L and CHF-CB30S choledochofiberscopes are only 2.8mm in outer diameter allowing easy insertion into the cystic and common bile duct. The CHF-CB30L and CHF-CB30S choledochofiberscopes are only 2.8mm in outer diameter allowing easy insertion into the cystic and common bile duct - The CHF-CB30 is available in two separate working lengths to meet all surgical needs. Some surgeons like to control and manipulate the endoscope by twisting the insertion tube and prefer the longer, L-type instrument, while others prefer using the control handle of the endoscope and use the S-type instrument. Large working channel 1.2mm - Despite the narrow diameter of the insertion tube, the CHF-CB30’s has a 1.2mm channel which facilitates distention and the use of a wide range of 3Fr instruments.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: The manual or automated leak test fails; bubbles from the bending section, insertion tube, or a channel port.
Root cause: A pinhole in the bending-rubber, a channel breach, or a damaged distal cap - fluid invasion risk.
Symptoms: Blurred, dark, spotty, half-image or no image; colour cast that will not white-balance.
Root cause: Fluid invasion into the CCD / fibre bundle, a broken image bundle, or lens-surface damage / adhesive fog.
Symptoms: Up/down or left/right angulation is loose, will not hold, over-tight, or limited range.
Root cause: Stretched or broken angulation wires, a worn brake, or bending-section damage.
Symptoms: Poor suction, no or weak air / water, will not pass a cleaning brush.
Root cause: Debris or a kink in the channel, a worn air/water or suction valve seat, or a crushed channel.
Symptoms: Dim or no illumination from the scope with a known-good light source.
Root cause: Broken light-guide fibres or a damaged light-guide connector.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Bending-rubber / distal cap | Frequent wear item | $$ | Moderate |
| Angulation wires / brake assembly | Periodic replacement | $$$ | Difficult |
| Air/water & suction valves / seats | Frequent wear item | $ | Easy |
| Image bundle / CCD repair (OEM/3rd-party) | Periodic replacement | $$$$ | Expert-level |
| Light-guide bundle / connector | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.