Nellcor · Model Puritan Bennett NPB 500
Puritan Bennett NPB-500 Simplicity Spirometer The Simplicity Spirometer System is the simple and affordable solution for diagnostic and screening spirometry. The system features the NPB-500, a compact, hand-held spirometer that offers true portability and is as quick and easy to use as its name. The system performs all of the basic spirometry measurements, and includes automated comparisons to predicted values, pre-post medication comparison, interpretation from test results, and comprehensive testing quality control. They system meets American Thoracic Athletics Society standards for accuracy and precision, and meets National Lung Health Education Program (NLHEP) recommendation for screening spirometry. The system uses the new Nellcor Puritan Bennett FSII Flow Sensor, a low cost, disposable, single-patient-use pneumotach that minimizes cross contamination and fits both adulthood and pediatric patients.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: 3 L syringe check outside tolerance, volume / flow drift over the day, biological control out of range.
Root cause: Flow-sensor contamination or wear (pneumotach / turbine / ultrasonic), a leak, or ambient (BTPS) correction not set.
Symptoms: Erratic traces, poor repeatability, back-extrapolated volume errors.
Root cause: Worn or wet flow sensor, a damaged filter seat, or a leak in the tubing / mouthpiece connection.
Symptoms: DLCO or FRC results implausible, gas-analyser calibration fails, drift.
Root cause: CO / CH4 / He / tracer analyser drift or an analyser fault, or a demand-valve / gas-supply issue.
Symptoms: Box pressure will not stabilise, thermal drift, door-seal leak.
Root cause: Door seal, the box pressure transducer, or thermal equilibration.
Symptoms: Test hangs, results will not save, print or export.
Root cause: Workstation PC, software, or the printer.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Calibrate / verify the syringe periodically.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow sensor (pneumotach / turbine / ultrasonic) | Periodic replacement | $$ | Fairly easy |
| Calibration syringe (3 L) Calibrate / verify the syringe periodically. | Major service part | $$ | Easy |
| Gas analyser cell (DLCO / lung-volume systems) | Periodic replacement | $$$ | Moderate |
| Plethysmograph door seal / pressure transducer | Periodic replacement | $$ | Moderate |
| Sample lines / demand valve | Frequent wear item | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
A software issue may lead to ventilator inoperative situations.
Covidien is recalling certain Puritan Bennett 980 Ventilator Systems due to a blacklight inverter PCBA where a capacitor component can fail, causing a dim but functional screen and thermal odor.
In the case of a loss of GUI display due to a Backlight Inverter PCBA failure, the ventilator continues to provide uninterrupted ventilatory support at the programmed settings for the patient. However, there is a loss of display and thus there is a necessity to move the patient to another ventilator.
Covidien is conducting a medical device field correction for specific Puritan Bennett 840 Ventilators due to customer reports of loss of graphical user interface (GUI) display information while the ventilator continues to provide breath support.
Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator systems. The Operators Manual for PB840 states a nominal life of one year and the Addendum incorrectly states 2 years.
Inflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
In rare instances, the safety valve may remain open after an occlusion is resolved. If the safety valve does not reset, the ventilator will detect a circuit disconnect condition. When this happens, the ventilator will annunciate a high-urgency audible alarm and the ventilator will not resume normal operation.
Wires on battery may loosen causing disconnection.
The internal speaker components are failing and do not produce an audio alarm in the event of patient emergency.