Three display options: plethysmographic waveform, large numerics and customizable real-time trend. On-screen quick guide provides instant help to assist caregivers with monitor operation. Convenient selection of adult or neonate alarm limits, and user-configurable power-on settings. User-selectable fast-averaging mode (2-3 seconds) to more precisely track rapidly changing saturations. On-screen viewing of 24-48 hours of SpO2 and pulse rate trends taken at 2-4 second intervals; printing capability. Variable pitch beep tone enables clinicians to hear changes in SpO2. Backup speaker with escalating alarm tones. Interfaces with certain multiparameter monitors; real-time patient information and true physiologic waveforms are displayed on host system. Works with the Oxinet™ III remote respiratory monitoring system to provide an easy-to-use, cost-effective monitoring solution for the general care floor. 7-hour battery life with battery gauge for convenient portability.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No saturation / pulse value, low signal-quality / "pulse search", implausible values.
Root cause: Worn or wrong SpO2 sensor, a damaged extension cable, poor site / perfusion, or motion.
Symptoms: Reading drops when the cable is moved, sensor not recognised.
Root cause: Broken conductors in the sensor / extension cable or a worn connector.
Symptoms: SpO2 consistently biased vs an arterial co-oximetry reference.
Root cause: Sensor degradation, or a monitor calibration / module fault.
Symptoms: No or quiet audible alarm.
Root cause: Speaker or audio-board fault - a safety-critical failure.
Symptoms: Short battery run time, will not power on.
Root cause: Aged battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| SpO2 sensor (reusable / disposable) | Frequent wear item | $$ | Easy |
| SpO2 extension cable | Frequent wear item | $$ | Easy |
| Battery | Periodic replacement | $$ | Fairly easy |
| Alarm speaker / audio board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.