Nellcor · Model N395
Simple, intuitive operation and easy-to-see display Compatible with the complete family of OxiMax Pulse Oximetry Sensors. Ideal for any clinical environment - monitors neonates through adults. Convenient selection of adult or neonate alarm limits, and user-configurable power-on settings. Works with the Oxinet III Central Station and Paging System to provide an easy-to-use, cost-effective monitoring solution for the General Care Floor. Compatible with Nellcor’s Score Analysis Software for overnight oximetry studies. User-selectable fast-averaging mode (2-3 seconds) to more precisely track rapidly changing saturations. On-screen viewing of 48 hours of SpO 2 and pulse rate trends taken at 4-second intervals; printing capability. Variable pitch beep tone enables clinicians to hear changes in SpO 2. Interfaces with certain multiparameter monitors; real-time patient information and true physiologic waveforms are displayed on host system.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No saturation / pulse value, "pulse search", low signal-quality, implausible values.
Root cause: Worn or wrong SpO2 sensor, a damaged pulse-oximetry cable, poor site / perfusion, or motion.
Symptoms: Reading drops when the cable is moved, sensor not recognised.
Root cause: Broken conductors in the sensor / DOC-10 cable or a worn connector.
Symptoms: SpO2 consistently biased vs an arterial reference.
Root cause: Sensor degradation, or a monitor calibration / module fault.
Symptoms: No or quiet audible alarm.
Root cause: Speaker or audio-board fault - a safety-critical failure.
Symptoms: Short battery run time, will not power on.
Root cause: Aged battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| SpO2 sensor (reusable / disposable) | Frequent wear item | $$ | Easy |
| Pulse-oximetry extension cable | Frequent wear item | $$ | Easy |
| Battery | Periodic replacement | $$ | Fairly easy |
| Alarm speaker / audio board | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.