Miramar Labs · Model MiraDry MD4000-MC
Non-invasive system to reduce sweat. The miraDry System MD4000 is indicated for use in the treatment of primary axillary hyperhidrosis plus unwanted underarm hair removal, and permanent reduction of underarm hair of all colors for Fitzpatrick skin types I – IV. Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. When used with the treatment of primary axillary hyperhidrosis, the miraDry System may reduce underarm odor.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Treatment ineffective, energy not reaching the set level, RF / microwave fault.
Root cause: Magnetron / microwave-source aging, a damaged handpiece / cable, or a generator fault.
Symptoms: Handpiece will not acquire or hold the skin, vacuum alarm.
Root cause: Vacuum pump, a leak in the handpiece / tubing, or a worn bioTip.
Symptoms: Skin cooling not delivered, over-temperature alarm, treatment aborts.
Root cause: Coolant supply / pump, tubing, or a temperature-sensor fault.
Symptoms: Console will not enable the handpiece, bioTip error.
Root cause: Handpiece connector wear or a bioTip fault / count exhausted.
Symptoms: Unresponsive display, will not power on.
Root cause: Power supply or control board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Microwave source / generator components | Periodic replacement | $$$$ | Expert-level |
| Reusable handpiece / cable | Periodic replacement | $$$ | Fairly easy |
| Single-use bioTips | Consumable | $$ | Easy |
| Vacuum pump | Periodic replacement | $$$ | Difficult |
| Coolant pump / tubing | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Update to labeling for the miraDry Console to remind users to only use approved lubricants in a thin layer to prevent potential burns and blisters.
Reports of skin burns on axilla after treatment with the miraDry System.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio