Mindray · Model BeneVision N1
Compact, lightweight (2.0 lbs) 5.5″ capacitive touchscreen display with smartphone-like, multi-gesture control Innovative design eliminates difficult-to-clean seams and gaps Consistent intuitive user interface across N-Series platform Supports display of up to 6 waveforms Comprehensive standard measurements include 3, 5, 6-lead ECG, Masimo RD SET® SpO2, 2 IBP, NIBP, respiration, and 2 temperatures Arrhythmia analysis and atrial fibrillation (Afib) detection. Optional ST analysis and QT/QTc interval monitoring. Extensive data storage plus up to 48 hours of full disclosure Serves as the core parameter module for N-Series monitors Flexible transport solution with optional integrated or modular CO2 IP44 grade particle and water resistant 1.2 m drop test endurance Multiple mounting solutions including transport module rack with bedrail hooks Standard 2.4/5 GHz dual band wireless communication Defibrillator synchronization Battery run time of 8 hours (without integrated CO2)
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, SpO2, temperature, CO2 or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a parameter-module / input-board fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: The host does not display a plugged-in module or a built-in parameter.
Root cause: Docking / connector fault or an internal board fault.
Symptoms: Blank, dim or distorted display; unresponsive touchscreen.
Root cause: Display, backlight, or controller fault.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack or a power-supply fault.
Symptoms: Monitor not shown at the central station, admit fails, intermittent drop.
Root cause: Wired / Wi-Fi network configuration or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| Parameter module / input board | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Display / touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.