MEDTRONIC MINIMED, INC. · Model MMT-751PNAH
PUMP MMT-751PNAH PRDGM 3.0 1.1F PK US PW
Intended use: A portable electronic device designed to provide continuous or intermittent subcutaneous delivery of insulin to a patient with diabetes mellitus, incorporating software to receive glucose status data from a monitor/transmitter and to determine the delivery of insulin based on pre-set thresholds using a dichotomous or binary algorithm. It may be badge-like, designed to adhere to skin or clothing or worn in a case attached to the belt or carried in a pocket and includes a reservoir, a pumping mechanism, and an infusion-rate controller. It may also allow additional manual administration/infusion and functionality to send data to a remote reporting system.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or hyperglycemia due to inaccuracies in insulin delivery (either under-delivery, over-delivery, or cessation of delivery).
A software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery
The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.
Medtronic MiniMed is recalling the MiniMed 620G and 640G insulin pumps because there are certain scenarios where the set Bolus screen will not timeout, which could cause confusion by showing a bolus amount that is no longer appropriate.
The firm is informing customers of a software error that might result in an inaccurate display of the Temp Basal time and duration in CareLink Clinical reports with data uploaded from the MiniMed 620G or 640G insulin pump.
Medtronic MiniMed is recalling the Implantable Insulin Pump because there have been reports of a pump malfunction resulting in the disconnection of the delivery of insulin prior to the end of the 7 year useful life of the device.
Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.