MEDTRONIC, INC. · Model 383069
LEAD 383069 SELECTSECURE US MRI
Intended use: An implantable wire with an electrode, insulated with non-conductive material except at its ends, which serves as an electrical conductor to transmit pacing impulses from an implanted pulse generator to the heart; it is not intended to conduct defibrillation impulses. It may transmit intrinsic electrical activity from the heart back to the pulse generator. The electrode end is normally placed through a vein to contact the endocardium where it is fixed, or to penetrate the myocardium of the interventricular septum adjacent to the HIS-Purkinje system; the other end is connected to the pulse generator. It may be impregnated with a steroid intended to reduce inflammation.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Identified devices may experience sudden battery power failure without alarm, power supply power-on-resets, and interruption of therapy. Potential patient harms include insufficient pacing support leading to cardiac arrest, syncope, or hospitalization; or system revision (explant/replacement) leading to infection, hematoma, or temporary impairment.
During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.