Maxtec · Model Tecotherm Neo
3 modes of operation for maximum flexibility in use Clinician can set and adjust parameters at the start and throughout treatment Data storage for allowing simple analysis of temperature profiles USB output to CSV Large colour screen is easy to see across the room Small / Lightweight portable device Internal battery restarts from same settings in event of power failure Smooth, linear rewarming Compatible with reusable or disposable accessories Optional use of skin probes Maintains normothermia after rewarming.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Servo or manual mode does not achieve the target rectal / skin temperature, slow response.
Root cause: Circulating fluid low or fouled, a blanket / hose kink or disconnect, a heater / chiller (Peltier) fault, or a temperature-probe fault.
Symptoms: Fluid-level or flow alarm, pump runs dry, therapy pauses.
Root cause: Sterile water below level (needs replacing every 2 months), an air lock, or a pump fault.
Symptoms: Implausible temperature, probe-fault alarm, servo mode unavailable.
Root cause: Damaged rectal / skin probe or its connector.
Symptoms: Fluid loss, damp blanket, frequent top-ups.
Root cause: Perforated blanket or worn quick-connect couplings.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Replace fluid and cleanse the system every 2 months.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Circulating sterile water + Chlorine Dioxide cleansing tablet Replace fluid and cleanse the system every 2 months. | Consumable | $ | Fairly easy |
| Temperature probes (rectal / skin) | Frequent wear item | $$ | Easy |
| Cooling / heating (Peltier) module | Major service part | $$$ | Difficult |
| Circulation pump | Major service part | $$ | Moderate |
| Patient blanket / wrap and hoses | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.