Maquet Critical Care AB · Model 6886611
Vaporizer Sevoflurane, QUICK FILL
Intended use: An assembly of electronic devices used for general inhalation anaesthesia (i.e., the delivery of oxygen (O2), nitrous oxide (N2O), other medical gases, and anaesthetic agents) and possibly total intravenous anaesthesia and local anaesthesia. Its main components are high, intermediate, and low pressure gas delivery systems, a breathing circuit (typically with a ventilator), and an anaesthetic gas scavenging system (AGSS). It has built-in alarms, analysers, control/monitoring systems, and typically has integrated circuitry/touchscreen displays for target controlled anaesthesia (TCA) for any type of patient (e.g., neonate, adult).
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Delivered agent concentration off, low/high agent alarm.
Root cause: Worn bellows or O-ring; agent contamination.
Symptoms: Low minute volume, volume loss, leak alarm.
Root cause: Loose connections, torn bellows, cracked hose.
Symptoms: High scavenger pressure, agent smell in OR.
Root cause: Obstructed scavenger line or failed regulator.
Symptoms: Rising EtCO2, patient rebreathing.
Root cause: CO2 absorbent saturated.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable, replace on schedule.
Requires periodic calibration.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| CO2 absorber canister Consumable, replace on schedule. | Consumable | $ | Easy |
| Scavenger interface & hose | Periodic replacement | $$ | Fairly easy |
| O2 pressure regulator | Periodic replacement | $$$ | Difficult |
| Vaporizer (agent-specific) Requires periodic calibration. | Major service part | $$$$ | Expert-level |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →